Literature DB >> 15742307

Plasma leakage of oxygenators in ECMO depends on the type of oxygenator and on patient variables.

B Meyns1, L Vercaemst, E Vandezande, H Bollen, D Vlasselaers.   

Abstract

This study was conducted to identify the causes of plasma leakage of oxygenators in extra corporeal membrane oxygenation (ECMO). From 1996 through 2002, 91 oxygenators were used in 62 patients undergoing ECMO for respiratory and/or cardiac failure. Several types of oxygenators were used (Medtronic Maxima, Minima, PRF, Medos Hilite). Patient variables and variables related to the ECMO set-up were analysed for their relationship with oxygenator failure by a time related multiple regression analysis (Cox). Oxygenator failure occurred in 26% of the cases. The analysis identified the type of oxygenator (p=0.0016), younger patient age (p=0.04) and the number of oxygenators used (p=0.03) as the independent significant risk factors. The type of oxygenator used has the most overwhelming effect (significantly less leakage with the Medos Hilite). In conclusion, leakage of oxygenators is predominantly caused by the type of oxygenator used. Patient variables (younger age and the number of oxygenators used in one patient) are also significant and allude to an inflammatory process as underlying mechanism of plasma leakage.

Entities:  

Mesh:

Year:  2005        PMID: 15742307     DOI: 10.1177/039139880502800106

Source DB:  PubMed          Journal:  Int J Artif Organs        ISSN: 0391-3988            Impact factor:   1.595


  4 in total

1.  Rapid-onset plasma leakage of extracorporeal oxygenation membranes possibly due to hyperbilirubinemia.

Authors:  Yoshiko Kida; Shinichiro Ohshimo; Michihito Kyo; Yuko Tanabe; Kei Suzuki; Koji Hosokawa; Nobuaki Shime
Journal:  J Artif Organs       Date:  2018-06-02       Impact factor: 1.731

2.  Gas Permeability and Mechanical Properties of Polyurethane-Based Membranes for Blood Oxygenators.

Authors:  Inês Coelho; Rita F Pires; Sérgio B Gonçalves; Vasco D B Bonifácio; Mónica Faria
Journal:  Membranes (Basel)       Date:  2022-08-24

3.  Towards Biohybrid Lung Development-Fibronectin-Coating Bestows Hemocompatibility of Gas Exchange Hollow Fiber Membranes by Improving Flow-Resistant Endothelialization.

Authors:  Michael Pflaum; Sophie Jurmann; Katherina Katsirntaki; Marisa Mälzer; Axel Haverich; Bettina Wiegmann
Journal:  Membranes (Basel)       Date:  2021-12-27

4.  Insights into Gradient and Anisotropic Pore Structures of Capiox® Gas Exchange Membranes for ECMO: Theoretically Verifying SARS-CoV-2 Permeability.

Authors:  Makoto Fukuda; Ryo Tanaka; Kazunori Sadano; Asako Tokumine; Tomohiro Mori; Hitoshi Saomoto; Kiyotaka Sakai
Journal:  Membranes (Basel)       Date:  2022-03-10
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.