Performance of a block detector PET scanner in imaging non-pure positron emitters--modelling and experimental validation with 124I.

S Robinson1, P J Julyan, D L Hastings, J Zweit.   

Abstract

The key performance measures of resolution, count rate, sensitivity and scatter fraction are predicted for a dedicated BGO block detector patient PET scanner (GE Advance) in 2D mode for imaging with the non-pure positron-emitting radionuclides 124I, 55Co, 61Cu, 62Cu, 64Cu and 76Br. Model calculations including parameters of the scanner, decay characteristics of the radionuclides and measured parameters in imaging the pure positron-emitter 18F are used to predict performance according to the National Electrical Manufacturers Association (NEMA) NU 2-1994 criteria. Predictions are tested with measurements made using 124I and show that, in comparison with 18F, resolution degrades by 1.2 mm radially and tangentially throughout the field-of-view (prediction: 1.2 mm), count-rate performance reduces considerably and in close accordance with calculations, sensitivity decreases to 23.4% of that with 18F (prediction: 22.9%) and measured scatter fraction increases from 10.0% to 14.5% (prediction: 14.7%). Model predictions are expected to be equally accurate for other radionuclides and may be extended to similar scanners. Although performance is worse with 124I than 18F, imaging is not precluded in 2D mode. The viability of 124I imaging and performance in a clinical context compared with 18F is illustrated with images of a patient with recurrent thyroid cancer acquired using both [124I]-sodium iodide and [18F]-2-fluoro-2-deoxyglucose.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15724539     DOI: 10.1088/0031-9155/49/24/008

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  6 in total

1.  Assessment of the short-lived non-pure positron-emitting nuclide (120)I for PET imaging.

Authors:  H Herzog; S M Qaim; L Tellmann; S Spellerberg; D Kruecker; H H Coenen
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-07-15       Impact factor: 9.236

2.  ARRONAX, a high-energy and high-intensity cyclotron for nuclear medicine.

Authors:  Ferid Haddad; Ludovic Ferrer; Arnaud Guertin; Thomas Carlier; Nathalie Michel; Jacques Barbet; Jean-François Chatal
Journal:  Eur J Nucl Med Mol Imaging       Date:  2008-05-09       Impact factor: 9.236

3.  Effect of the positron range of 18F, 68Ga and 124I on PET/CT in lung-equivalent materials.

Authors:  Gerrit J Kemerink; Mariëlle G W Visser; Renee Franssen; Emiel Beijer; Mariangela Zamburlini; Servé G E A Halders; Boudewijn Brans; Felix M Mottaghy; Gerrit J J Teule
Journal:  Eur J Nucl Med Mol Imaging       Date:  2011-02-02       Impact factor: 9.236

4.  A comparison of PET imaging characteristics of various copper radioisotopes.

Authors:  Heather Ann Williams; Simon Robinson; Peter Julyan; Jamal Zweit; David Hastings
Journal:  Eur J Nucl Med Mol Imaging       Date:  2005-10-29       Impact factor: 9.236

5.  Correction technique for cascade gammas in I-124 imaging on a fully-3D, Time-of-Flight PET Scanner.

Authors:  Suleman Surti; Ryan Scheuermann; Joel S Karp
Journal:  IEEE Trans Nucl Sci       Date:  2009-06       Impact factor: 1.679

6.  Iodine-124 PET dosimetry in differentiated thyroid cancer: recovery coefficient in 2D and 3D modes for PET(/CT) systems.

Authors:  Walter Jentzen; Reiner Weise; Jürgen Kupferschläger; Lutz Freudenberg; Wolfgang Brandau; Ronald Bares; Wolfgang Burchert; Andreas Bockisch
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-10-10       Impact factor: 9.236

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.