| Literature DB >> 15724233 |
Zoltán Kele1, Györgyi Ferenc, Eva Klement, Gábor K Tóth, Tamás Janáky.
Abstract
A simple sheathless capillary electrophoresis (CE)/mass spectrometry (MS) interface was constructed by combining widely used nanospray needles with fused-silica capillaries and it was successfully applied for the separation of peptides. The end of the CE capillary was pulled to a taper, etched and then fitted into the metal-coated nanospray borosilicate capillary. The nanospray needle can be used for several CE runs, but it can be easily and rapidly changed in the case of accidental breakage or evaporation of the coating. A fast capillary electrochromatographic method was also developed for MS analysis of peptides containing numerous basic amino acids. Copyright 2005 John Wiley & Sons, Ltd.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15724233 DOI: 10.1002/rcm.1866
Source DB: PubMed Journal: Rapid Commun Mass Spectrom ISSN: 0951-4198 Impact factor: 2.419