Quantitative analysis of patient-specific dosimetric IMRT verification.

G J Budgell1, B A Perrin, J H L Mott, J Fairfoul, R I Mackay.   

Abstract

Patient-specific dosimetric verification methods for IMRT treatments are variable, time-consuming and frequently qualitative, preventing evidence-based reduction in the amount of verification performed. This paper addresses some of these issues by applying a quantitative analysis parameter to the dosimetric verification procedure. Film measurements in different planes were acquired for a series of ten IMRT prostate patients, analysed using the quantitative parameter, and compared to determine the most suitable verification plane. Film and ion chamber verification results for 61 patients were analysed to determine long-term accuracy, reproducibility and stability of the planning and delivery system. The reproducibility of the measurement and analysis system was also studied. The results show that verification results are strongly dependent on the plane chosen, with the coronal plane particularly insensitive to delivery error. Unexpectedly, no correlation could be found between the levels of error in different verification planes. Longer term verification results showed consistent patterns which suggest that the amount of patient-specific verification can be safely reduced, provided proper caution is exercised: an evidence-based model for such reduction is proposed. It is concluded that dose/distance to agreement (e.g., 3%/3 mm) should be used as a criterion of acceptability. Quantitative parameters calculated for a given criterion of acceptability should be adopted in conjunction with displays that show where discrepancies occur. Planning and delivery systems which cannot meet the required standards of accuracy, reproducibility and stability to reduce verification will not be accepted by the radiotherapy community.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15715426     DOI: 10.1088/0031-9155/50/1/009

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  Practical guidelines for routine intensity-modulated radiotherapy verification: pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry.

Authors:  A J Vinall; A J Williams; V E Currie; A Van Esch; D Huyskens
Journal:  Br J Radiol       Date:  2010-11       Impact factor: 3.039

2.  Implementation and validation of a commercial portal dosimetry software for intensity-modulated radiation therapy pre-treatment verification.

Authors:  C Varatharaj; Eugenia Moretti; M Ravikumar; Maria Rosa Malisan; Sanjay S Supe; Renato Padovani
Journal:  J Med Phys       Date:  2010-10

3.  Portal dosimetry for pretreatment verification of IMRT plan: a comparison with 2D ion chamber array.

Authors:  Dayananda Shamurailatpam Sharma; Vaibav Mhatre; Malhotra Heigrujam; Kaustav Talapatra; Suman Mallik
Journal:  J Appl Clin Med Phys       Date:  2010-08-19       Impact factor: 2.102

4.  Clinical implementation and error sensitivity of a 3D quality assurance protocol for prostate and thoracic IMRT.

Authors:  Gueorgui Gueorguiev; Christopher Cotter; Julie Catherine Turcotte; Bruce Crawford; Gregory Sharp; Mufeed Mah'D
Journal:  J Appl Clin Med Phys       Date:  2015-09-08       Impact factor: 2.102

5.  Clinical Experience of Intensity Modulated Radiotherapy Pre-Treatment Quality Assurance for Carcinoma Head and Neck Patients with EPID and IMatriXX in Rural Center.

Authors:  More M; Jain V; Gurjar O P
Journal:  J Biomed Phys Eng       Date:  2020-12-01

6.  A comparative analysis of Acuros XB and the analytical anisotropic algorithm for volumetric modulation arc therapy.

Authors:  Raju P Srivastava; K Basta; Werner De Gersem; Carlos De Wagter
Journal:  Rep Pract Oncol Radiother       Date:  2021-06-09

7.  Effect of contrast enhanced CT scans on heterogeneity corrected dose computations in the lung.

Authors:  Nichola A Burridge; Carl G Rowbottom; Paul A Burt
Journal:  J Appl Clin Med Phys       Date:  2006-11-28       Impact factor: 2.102

8.  A simple model for predicting the signal for a head-mounted transmission chamber system, allowing IMRT in-vivo dosimetry without pretreatment linac time.

Authors:  Daniel Johnson; Steven J Weston; Vivian P Cosgrove; David I Thwaites
Journal:  J Appl Clin Med Phys       Date:  2014-07-08       Impact factor: 2.102

9.  Practical application of Octavius® -4D: Characteristics and criticalities for IMRT and VMAT verification.

Authors:  Patrizia Urso; Rita Lorusso; Luca Marzoli; Daniela Corletto; Paolo Imperiale; Annalisa Pepe; Lorenzo Bianchi
Journal:  J Appl Clin Med Phys       Date:  2018-07-16       Impact factor: 2.102

10.  Patient-Specific Quality Assurance Protocol for Volumetric Modulated Arc Therapy using Dose Volume Histogram.

Authors:  Christopher Low; Warren Toye; Peter Phung; Christopher Huston
Journal:  J Med Phys       Date:  2018 Apr-Jun