What is the adequate sampling interval of the ECG signal for heart rate variability analysis in the time domain?

Laszlo Hejjel1, Elizabeth Roth.   

Abstract

Heart rate variability (HRV) analysis is considered a simple method for investigating neurocardiac regulation. However, measures of HRV may be corrupted by technical artifacts. In order to investigate the consequences of digitization errors on the time domain parameters, HRV measures from model tachograms resampled at different rates were compared. Two 375-element tachograms from human ECG tracings were shifted to a mean of 800 ms and stretched to standard deviations (SD) of 5-120 ms in 5 ms steps. All were resampled at 1-10 ms in 1 ms steps, 10 times repetitively at each interval. The mean, SD, relative accuracy error (RAE) and relative precision error (RPE) were calculated from the mean RR-interval, SD (SDNN), root mean square of successive RR-differences (RMSSD) and the percentage of consecutive RR-interval differences greater than 50 ms (pNN50). The RAE and RPE of the mean heart rate were below 0.1%. In the series with 5 ms SD, the SDNN-RAE exceeded 30% at 10 ms SI, its RPE was lower than 2% all through. Resampling the 15 ms SD tachogram at 1, 2, 4, or 10 ms resulted in RMSSD-RAE of 0.7%, 2.5%, 7.8% and 45.1%, respectively, while its RPE remained below 5%. The pNN50 shows poor accuracy and precision. An ECG sampling interval of 1 ms is recommended for HRV analysis without interpolation in order to get accurate time domain measures even in seriously reduced-variability samples. However, a lower sampling rate may be satisfactory in cases where higher variability is expected.

Entities:  

Mesh:

Year:  2004        PMID: 15712719     DOI: 10.1088/0967-3334/25/6/006

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  10 in total

1.  Test-Retest Reliability of Pediatric Heart Rate Variability: A Meta-Analysis.

Authors:  Oren M Weiner; Jennifer J McGrath
Journal:  J Psychophysiol       Date:  2016-07-27       Impact factor: 1.333

2.  Transcutaneous Carbon Dioxide Treatment Affects Heart Rate Variability - A Pilot Study.

Authors:  Zita Kreska; Balázs Németh; István Kiss; Iván Péter; Zénó Ajtay; László Hejjel
Journal:  In Vivo       Date:  2018 Sep-Oct       Impact factor: 2.155

3.  Design and implementation of an ultra-low energy FFT ASIC for processing ECG in Cardiac Pacemakers.

Authors:  Safwat Mostafa; Eugene B John; Manoj M Panday
Journal:  IEEE Trans Very Large Scale Integr VLSI Syst       Date:  2018-12-14       Impact factor: 2.312

4.  A novel handheld device for use in remote patient monitoring of heart failure patients--design and preliminary validation on healthy subjects.

Authors:  Luca Pollonini; Nithin O Rajan; Shuai Xu; Sridhar Madala; Clifford C Dacso
Journal:  J Med Syst       Date:  2010-06-08       Impact factor: 4.460

5.  Physical Vascular Therapy (BEMER) Affects Heart Rate Asymmetry in Patients With Coronary Heart Disease.

Authors:  Zita Kreska; Péter Mátrai; Balázs Nemeth; Bella Ajtay; István Kiss; László Hejjel; Zénó Ajtay
Journal:  In Vivo       Date:  2022 May-Jun       Impact factor: 2.406

6.  Influence of ECG sampling frequency on spectral analysis of RR intervals and baroreflex sensitivity using the EUROBAVAR data set.

Authors:  Tjalf Ziemssen; Julia Gasch; Heinz Ruediger
Journal:  J Clin Monit Comput       Date:  2008-03-27       Impact factor: 2.502

7.  The acute effects of joint manipulative techniques on markers of autonomic nervous system activity: a systematic review and meta-analysis of randomized sham-controlled trials.

Authors:  Mathieu Picchiottino; Charlotte Leboeuf-Yde; Olivier Gagey; David M Hallman
Journal:  Chiropr Man Therap       Date:  2019-03-12

8.  Optimizing sampling rate of wrist-worn optical sensors for physiologic monitoring.

Authors:  Brinnae Bent; Jessilyn P Dunn
Journal:  J Clin Transl Sci       Date:  2020-08-25

9.  Impact of observational error on heart rate variability analysis.

Authors:  Monika Petelczyc; Jan Jakub Gierałtowski; Barbara Żogała-Siudem; Grzegorz Siudem
Journal:  Heliyon       Date:  2020-05-19

10.  Enhancing the Robustness of Smartphone Photoplethysmography: A Signal Quality Index Approach.

Authors:  Ivan Liu; Shiguang Ni; Kaiping Peng
Journal:  Sensors (Basel)       Date:  2020-03-30       Impact factor: 3.576

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.