Detection of hypoventilation during thoracoscopy: combined cutaneous carbon dioxide tension and oximetry monitoring with a new digital sensor.

Prashant N Chhajed1, Bruno Kaegi, Rajeevan Rajasekaran, Michael Tamm.   

Abstract

BACKGROUND: Changes in Paco(2) have not been described during thoracoscopy under sedation-assisted local anesthesia. We hypothesized that hypoventilation might occur secondary to administration of sedatives and decreased ventilation in one lung. AIM: Prospectively measure cutaneous carbon dioxide tension (Pcco(2)) in addition to pulse oximetric saturation (Spo(2)) using a new combined digital sensor to examine the occurrence of hypoventilation during thoracoscopy under sedation-assisted local anesthesia.
SETTING: University hospital.
METHODS: Following validation studies, Pcco(2) was prospectively measured in 16 consecutive patients undergoing thoracoscopy under sedation-assisted local anesthesia using a combined digital earlobe sensor measuring Spo(2) (percentage) and Pcco(2) (millimeters of mercury). All patients received supplemental oxygen. Routine BP monitoring and Spo(2) was continued. Patients received IV hydrocodone, 5 mg, and intermittent boluses or IV midazolam and pethidine.
RESULTS: Mean baseline Pcco(2) measurement was 39.1 +/- 7.2 mm Hg (+/- SD) [range, 27.5 to 50.5 mm Hg], and peak measurement during the procedure was 52.3 +/- 10.3 mm Hg (range, 37.2 to 77 mm Hg) [p < 0.001]. Median and mean changes in Pcco(2) measurement from baseline were 13.0 mm Hg and 13.2 +/- 5.3 mm Hg (range, 5.5 to 27.8 mm Hg), respectively. Mean fall in Spo(2) during the procedure was 4.6 +/- 3.2% (range, 1 to 14%).
CONCLUSIONS: Thoracoscopy performed under sedation-assisted local anesthesia is associated with significant hypoventilation. Combined measurement of Spo(2) and Pcco(2) during thoracoscopy is a novel approach in the monitoring of ventilation, enhancing patient safety, and might allow to guide the administration of sedation in a better way.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15706000     DOI: 10.1378/chest.127.2.585

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  8 in total

1.  Clinical validation of a digital transcutaneous PCO2/SpO2 ear sensor in adult patients after cardiac surgery.

Authors:  Werner Baulig; Philipp Schütt; Hans R Roth; Josef Hayoz; Edith R Schmid
Journal:  J Clin Monit Comput       Date:  2007-08-16       Impact factor: 2.502

2.  Safety of carbon dioxide insufflation during gastric endoscopic submucosal dissection in patients with pulmonary dysfunction under conscious sedation.

Authors:  Jun Takada; Hiroshi Araki; Fumito Onogi; Takayuki Nakanishi; Masaya Kubota; Takashi Ibuka; Masahito Shimizu; Hisataka Moriwaki
Journal:  Surg Endosc       Date:  2014-10-16       Impact factor: 4.584

3.  Utility of Transcutaneous Capnography for Optimization of Non-Invasive Ventilation Pressures.

Authors:  Prashant N Chhajed; Simone Gehrer; Kamlesh V Pandey; Preyas J Vaidya; Joerg D Leuppi; Michael Tamm; Werner Strobel
Journal:  J Clin Diagn Res       Date:  2016-09-01

4.  Transcutaneous monitoring of partial pressure of carbon dioxide during endoscopic submucosal dissection of early colorectal neoplasia with carbon dioxide insufflation: a prospective study.

Authors:  Tsuyoshi Kikuchi; Kuang-I Fu; Yutaka Saito; Toshio Uraoka; Masakatsu Fukuzawa; Syusei Fukunaga; Taku Sakamoto; Takeshi Nakajima; Takahisa Matsuda
Journal:  Surg Endosc       Date:  2010-02-23       Impact factor: 4.584

5.  Carbon dioxide insufflation reduces residual gas in the gastrointestinal tract following colorectal endoscopic submucosal dissection.

Authors:  Tomohiko Sugiyama; Hiroshi Araki; Noritaka Ozawa; Jun Takada; Masaya Kubota; Takashi Ibuka; Masahito Shimizu
Journal:  Biomed Rep       Date:  2018-01-17

Review 6.  Medical thoracoscopy and its evolving role in the diagnosis and treatment of pleural disease.

Authors:  Vivek Murthy; Jamie L Bessich
Journal:  J Thorac Dis       Date:  2017-09       Impact factor: 2.895

7.  Transcutaneous partial pressure of carbon dioxide monitoring during EUS-guided drainage of peripancreatic fluid collections using carbon dioxide insufflation: A prospective study.

Authors:  Liu Xiang; Sun Jiayi; Wang Guoxin; Ge Nan; Wang Sheng; Guo Jintao; Sun Siyu
Journal:  Endosc Ultrasound       Date:  2020 Jan-Feb       Impact factor: 5.628

8.  Comparison of tracheal intubation with controlled ventilation and laryngeal mask airway with spontaneous ventilation for thoracoscopic bullectomy.

Authors:  Xiu-Liang Li; Xiang-Bo He; Lei Wan; Chun-Quan Liu; Yong Cui
Journal:  Medicine (Baltimore)       Date:  2020-04       Impact factor: 1.817

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.