Importance of dosimeter calibration method on nebulizer output.

David C Todd1, Beth E Davis, Audrey J Smycniuk, Donald W Cockcroft.   

Abstract

BACKGROUND: We have observed that dosimeter-run nebulizers have a much smaller output when manually activated than when breath activated; however, this has not been adequately investigated.
OBJECTIVE: To evaluate the effect of different calibration methods on nebulizer output.
METHODS: Six healthy subjects performed all calibrations. The nebulizers were operated by 2 different dosimeters and were calibrated to produce 9 microL per actuation by breath activation followed by exhalation to the room. The nebulizers were then operated at these identical settings, and the output determined in 3 ways: (1) breath activation followed by exhalation to the room, (2) breath activation with exhalation into the nebulizer, and (3) manual activation (with no subject using the nebulizer). These 3 methods were termed regular, rebreathe, and manual, respectively.
RESULTS: There was a large and statistically significant difference in nebulizer output among the 3 methods. The measured rebreathe outputs (5.6 and 5.7 microL per actuation) were approximately two thirds and the manual outputs (3.2 and 3.9 microL per actuation) were approximately one third of the regular calibration outputs (8.6 and 8.9 microL per actuation); the 2 values are for the 2 dosimeters. The results were highly statistically significant (P < .001).
CONCLUSIONS: The method by which a nebulizer-dosimeter system is calibrated results in different nebulizer outputs. This has a high likelihood of influencing the concentration of methacholine causing a 20% decrease in volume in the first second of forced expiration.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15702815     DOI: 10.1016/S1081-1206(10)61284-7

Source DB:  PubMed          Journal:  Ann Allergy Asthma Immunol        ISSN: 1081-1206            Impact factor:   6.347


  1 in total

1.  Airway resistance at maximum inhalation as a marker of asthma and airway hyperresponsiveness.

Authors:  Nancy T Mendonça; Jennifer Kenyon; Adam S LaPrad; Sohera N Syeda; George T O'Connor; Kenneth R Lutchen
Journal:  Respir Res       Date:  2011-07-15
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.