| Literature DB >> 15679147 |
Olga González-García1, Cristina Ariño, José Manuel Díaz-Cruz, Miquel Esteban.
Abstract
A voltammetric detection mode (VD) in conjunction with multivariate curve resolution with alternating least squares (MCR-ALS) method is applied to the analysis of cysteine-containing compounds and compared with a well established amperometric detection (AD) mode in a thin-layer dual Hg/Au cell. VD-MCR-ALS provides an increase in selectivity for cases where satisfactory separation of electroactive compounds is not allowed. However, concentrations needed for a good quantification in VD are higher than in AD due to much large contribution of background in VD.Entities:
Mesh:
Substances:
Year: 2005 PMID: 15679147 DOI: 10.1016/j.chroma.2004.11.018
Source DB: PubMed Journal: J Chromatogr A ISSN: 0021-9673 Impact factor: 4.759