OBJECTIVE: Our aim was to evaluate the performance of hypertonic saline (HS)-enhanced bipolar radiofrequency ablation using wet-cooled electrodes versus monopolar radiofrequency ablation to create coagulation necrosis in explanted bovine liver. CONCLUSION: HS-enhanced bipolar radiofrequency ablation using the wet-cooled electrodes shows better performance in creating coagulation necrosis than the monopolar mode.
OBJECTIVE: Our aim was to evaluate the performance of hypertonic saline (HS)-enhanced bipolar radiofrequency ablation using wet-cooled electrodes versus monopolar radiofrequency ablation to create coagulation necrosis in explanted bovine liver. CONCLUSION:HS-enhanced bipolar radiofrequency ablation using the wet-cooled electrodes shows better performance in creating coagulation necrosis than the monopolar mode.