| Literature DB >> 15663861 |
Christian Drosten1, Hans Wilhelm Doerr, Wilina Lim, Klaus Stöhr, Matthias Niedrig.
Abstract
Inactivated severe acute respiratory syndrome-associated coronavirus samples were used for an external quality assurance study within the World Health Organization SARS Reference and Verification Network and other reference institutions. Of 58 participants, 51 correctly detected virus in all samples > or =9,400 RNA copies per milliliter and none in negative samples. Commercial test kits significantly improved the outcome.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15663861 PMCID: PMC3323363 DOI: 10.3201/eid1012.040416
Source DB: PubMed Journal: Emerg Infect Dis ISSN: 1080-6040 Impact factor: 6.883
Positive samples in test panela
| Sample code | SARS-CoV strain | Virus RNA concentration copies/mL | Fraction of laboratories with positive detection (%) |
|---|---|---|---|
| S-CV2 | Frankfurt 1 | 940,000 | 100 |
| S-CV9 | Frankfurt 1 | 94,000 | 98.3 |
| S-CV6 | HKU-1 | 23,500 | 98.3 |
| S-CV4 | Frankfurt 1 | 9,400 | 94.8 |
| S-CV10 | HKU-1 | 2,350 | 87.9 |
| S-CV1 | Frankfurt 1 | 940 | 70.7 |
| S-CV5 | Frankfurt 1 | 94 | 43.1 |
aSARS-CoV, severe acute respiratory syndrome–associated coronavirus.
FigureProbit analysis of the fractions of laboratories achieving a positive result (y-axis) in relation to the virus RNA concentration in a given positive sample (x-axis). Data points represent individual samples in proficiency test panel. The thick line is the regression line calculated on the basis of a probit model (dose-response curve); the thin lines are 95% confidence intervals. Data fit into the model with p < 0.0001.
Factors influencing the performance of laboratoriesa
| Possible technical influence factors | No. of laboratories | Positive influence on sensitivity p value |
|---|---|---|
| Qiagen viral RNA extraction kit | 38 | 0.9 |
| Roche MagnaPure/HighPure extraction kit | 7 | 0.2 |
| Silica particle-based extraction method (Boom) | 9 | 0.9 |
| Primers originally developed in own laboratory | 16 | 0.5 |
| Any nested PCR assay | 25 | 0.9 |
| Any real-time PCR assay | 37 | 0.7 |
| Any commercial test kit | 14 | 0.03 |
aAnalysis of Variance (ANOVA) by factor, eliminating the influence of other factors; PCR, polymerase chain reaction.