[Potential advantages of the MRI for the radiofrequency ablation of liver tumors].

G Gaffke1, B Gebauer, M Gnauck, F D Knollmann, T Helmberger, J Ricke, H Oettle, R Felix, C Stroszczynski.   

Abstract

PURPOSE: To present first results of radiofrequency ablation of liver tumors using a new MR compatible applicator.
MATERIALS AND METHODS: We performed 37 interventions in 20 patients (mean age 58.6 years) with primary intrahepatic malignancies or metastases: colorectal carcinoma n = 6, hepatocellular carcinoma n = 3, pancreatic carcinoma n = 4, sarcoma n = 2, cholangiocellular carcinoma n = 1, carcinoma of the tonsil n = 1, breast carcinoma n = 1, gastric carcinoma n = 1, and gastrointestinal stroma tumor n = 1. Interventions were performed under CT-guidance with CT fluoroscopy (n = 32) and under MR-guidance (n = 5) using fast T1-weighted sequences in breath-hold technique. RFA was performed with the RF-generator (150 W) under local anesthesia and sedation using MR compatible applicators (Starburst XL, Rita Medical Systems, USA) together with the appropriate Soft Tissue Introducer System. Intra-interventional control was performed with intrahepatically or intralesionally placed introducer system or applicator. MRI was performed with plain breath-triggered T2-weighted turbo spin echo sequences (TSE T2) with fat saturation.
RESULTS: All interventions were performed without major events. The mean diameter of induced coagulation was 4.0 (+/- 0.7) cm. Repositioning was necessary in 8 interventions (21 %) after detection of residual tumor on an intra-interventional MRI. After a mean follow-up of 6.5 (+/- 1.2) months, the local tumor control rate was 92 %.
CONCLUSION: MR-compatible RF applicators offer the opportunity for intra-interventional detection of residual tumor during RF ablations by use of sensitive MRI sequences. These procedures may lead to a higher confidence in tumour ablation and may reduce the number of re-interventions and local recurrences of intrahepatic tumors.

Entities:  

Mesh:

Year:  2005        PMID: 15657824     DOI: 10.1055/s-2004-813643

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  2 in total

1.  Interaction between grounding pads used for RF ablation therapy and magnetic resonance imaging.

Authors:  Christina Schraml; Hansjörg Graf; Andreas Boss; Stephan Clasen; Martin Leibfritz; Philippe L Pereira; Claus D Claussen; Fritz Schick
Journal:  MAGMA       Date:  2005-12-13       Impact factor: 2.310

2.  MR-guided liver tumor ablation employing open high-field 1.0T MRI for image-guided brachytherapy.

Authors:  Jens Ricke; Markus Thormann; Matthias Ludewig; Kerstin Jungnickel; Oliver Grosser; Christian Wybranski; Nils Peters; Peter Hass; Jürgen Bunke; Frank Fischbach
Journal:  Eur Radiol       Date:  2010-03-20       Impact factor: 5.315

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.