Abstract
A fully 3D small PET scanner based on a novel detection principle for gamma rays is described. It uses BaF2 scintillator and photosensitive wire chambers. Extensive tests with technical prototypes have shown that such a system will have a detection efficiency for gamma rays comparable with what can be obtained with the more traditional approach, and a spatial resolution determined by the size of the crystals. The expected performances of the scanner, based on our measurements and on simulations, are given.Entities:
Mesh:
Year: 1992 PMID: 1565695 DOI: 10.1088/0031-9155/37/3/010
Source DB: PubMed Journal: Phys Med Biol ISSN: 0031-9155 Impact factor: 3.609