Literature DB >> 15649749

An object-oriented designed finite-difference time-domain simulator for electromagnetic analysis and design in MRI--applications to high field analyses.

Q Wei1, F Liu, L Xia, S Crozier.   

Abstract

This paper presents a finite-difference time-domain (FDTD) simulator for electromagnetic analysis and design applications in MRI. It is intended to be a complete FDTD model of an MRI system including all RF and low-frequency field generating units and electrical models of the patient. The program has been constructed in an object-oriented framework. The design procedure is detailed and the numerical solver has been verified against analytical solutions for simple cases and also applied to various field calculation problems. In particular, the simulator is demonstrated for inverse RF coil design, optimized source profile generation, and parallel imaging in high-frequency situations. The examples show new developments enabled by the simulator and demonstrate that the proposed FDTD framework can be used to analyze large-scale computational electromagnetic problems in modern MRI engineering.

Entities:  

Mesh:

Year:  2005        PMID: 15649749     DOI: 10.1016/j.jmr.2004.10.009

Source DB:  PubMed          Journal:  J Magn Reson        ISSN: 1090-7807            Impact factor:   2.229


  1 in total

Review 1.  Biochemical and physiological MR imaging of skeletal muscle at 7 tesla and above.

Authors:  Gregory Chang; Ligong Wang; Arturo Cárdenas-Blanco; Mark E Schweitzer; Michael P Recht; Ravinder R Regatte
Journal:  Semin Musculoskelet Radiol       Date:  2010-05-18       Impact factor: 1.777