An automated system for stable isotope and concentration analyses of CO2 from small atmospheric samples.

Andrew J Schauer1, Michael J Lott, Craig S Cook, James R Ehleringer.   

Abstract

We have developed an automated, continuous-flow isotope ratio mass spectrometry (CF-IRMS) system for the analysis of delta(13)C, delta(18)O, and CO(2) concentration (micromol mol(-1)) ([CO(2)]) from 2 mL of atmospheric air. Two replicate 1 mL aliquots of atmospheric air are sequentially sampled from fifteen 100 mL flasks. The atmospheric sample is inserted into a helium stream and sent through a gas chromatograph for separation of the gases and subsequent IRMS analysis. Two delta(13)C and delta(18)O standards and five [CO(2)] standards are run with each set of fifteen samples. We obtained a precision of 0.06 per thousand, 0.11 per thousand, and 0.48 micromol mol(-1) for delta(13)C, delta(18)O, and [CO(2)], respectively, by analyzing fifty 100 mL samples filled from five cylinders with a [CO(2)] range of 275 micromol mol(-1). Accuracy was determined by comparison with established methods (dual-inlet IRMS, and nondispersive infrared gas analysis) and found to have a mean offset of 0.00 per thousand, -0.09 per thousand, and -0.26 micromol mol(-1) for delta(13)C and delta(18)O, and [CO(2)], respectively. Copyright (c) 2005 John Wiley & Sons, Ltd.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15645505     DOI: 10.1002/rcm.1792

Source DB:  PubMed          Journal:  Rapid Commun Mass Spectrom        ISSN: 0951-4198            Impact factor:   2.419


  3 in total

1.  Inferring biogenic and anthropogenic carbon dioxide sources across an urban to rural gradient.

Authors:  D E Pataki; T Xu; Y Q Luo; J R Ehleringer
Journal:  Oecologia       Date:  2007-02-14       Impact factor: 3.298

2.  Use of a Dynamic Enclosure Approach to Test the Accuracy of the NDIR Sensor: Evaluation Based on the CO2 Equilibration Pattern.

Authors:  Sudhir Kumar Pandey; Ki-Hyun Kim; Sun-Ho Lee
Journal:  Sensors (Basel)       Date:  2007-12-20       Impact factor: 3.576

3.  Successive and automated stable isotope analysis of CO2 , CH4 and N2 O paving the way for unmanned aerial vehicle-based sampling.

Authors:  Simon Leitner; Rebecca Hood-Nowotny; Andrea Watzinger
Journal:  Rapid Commun Mass Spectrom       Date:  2020-12-30       Impact factor: 2.419

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.