Literature DB >> 1564461

Temperature influence in different orthopaedic saw blades.

S Toksvig-Larsen1, L Ryd, A Lindstrand.   

Abstract

Laboratory tests were carried out on ox bone to evaluate the thermal effect of eight different saw blades while cutting cortical bone. These saw blades represented the usual clinical blades as well as saw blades specially manufactured in an attempt to decrease the temperature. Temperatures between 34 degrees C and 450 degrees C were registered in the saw blades. Only three measurements (of 219 tests) were below 44 degrees-47 degrees C, which is a critical limit for heat-induced bone necrosis. This test indicates that alternating saw blade design is not a way to control the temperature elevation during cutting of bone in orthopaedic procedures.

Entities:  

Mesh:

Year:  1992        PMID: 1564461     DOI: 10.1016/0883-5403(92)90027-n

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  3 in total

1.  [Cup & cone reamers for arthrodesis of the first metatarsophalangeal joint].

Authors:  Hans-Peter Kundert
Journal:  Oper Orthop Traumatol       Date:  2010-10       Impact factor: 1.154

2.  New implant designs for fracture fixation in osteoporotic bone.

Authors:  J Goldhahn; J Seebeck; R Frei; B Frenz; I Antoniadis; E Schneider
Journal:  Osteoporos Int       Date:  2004-11-05       Impact factor: 4.507

3.  Accelerated bone repair after plasma laser corticotomies.

Authors:  Philipp Leucht; Kentson Lam; Jae-Beom Kim; Mark A Mackanos; Dmitrii M Simanovskii; Michael T Longaker; Christopher H Contag; H Alan Schwettman; Jill A Helms
Journal:  Ann Surg       Date:  2007-07       Impact factor: 12.969

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.