OBJECT: Spinal arthroplasty may become the next gold standard for the treatment of degenerative cervical spine disease. This new modality must be studied rigorously to ensure in vivo efficacy and safety. The authors review the preliminary clinical experience and radiographic outcomes following insertion of the Bryan Cervical Disc Prosthesis (Medtronic Sofamor Danek, Memphis, TN).
METHODS: This prospective cohort study included 26 patients undergoing single- or two-level implantation of the Bryan artificial cervical disc for treatment of cervical degenerative disc disease resulting in radiculopathy and/or myelopathy. Radiographic and clinical assessments were made preoperatively 1.5, 3 months, and at 6, 12, and up to 24 months postoperatively. The Neck Disability Index (NDI) and Short Form-36 (SF-36) questionnaires were used to assess pain and functional outcomes. Segmental sagittal rotation from C2-3 to C6-7 was measured using quantitative motion analysis software. A total of 30 Bryan discs were placed in 26 patients. A single-level procedure was performed in 22 patients and a two-level procedure in the other four. Follow-up duration ranged from 1.5 to 27 months, with a mean duration of 12.3 months. A statistically significant improvement in the mean NDI scores was seen between pre- and late postoperative follow-up evaluations. A trend toward improvement in the SF-36 physical component was also found. Motion was preserved in the treated spinal segments (mean range of motion 7.8 degrees ) for up to 24 months postsurgery. The relative contribution of each segment to overall spinal sagittal rotation differed depending on whether the disc was placed at C5-6 or C6-7. Overall cervical motion (C2-7) was moderately increased on late follow-up evaluations.
CONCLUSIONS: The Bryan artificial cervical disc provided in vivo functional spinal motion at the treated level. Overall cervical spinal motion was not significantly altered. Sagittal rotation did not change significantly at any level after surgery.

Entities:  

Mesh:

Year:  2004        PMID: 15636565     DOI: 10.3171/foc.2004.17.3.9

Source DB:  PubMed          Journal:  Neurosurg Focus        ISSN: 1092-0684            Impact factor:   4.047


  26 in total

Review 1.  Cervical spine alignment in disc arthroplasty: should we change our perspective?

Authors:  Alberto Di Martino; Rocco Papalia; Erika Albo; Leonardo Cortesi; Luca Denaro; Vincenzo Denaro
Journal:  Eur Spine J       Date:  2015-10-06       Impact factor: 3.134

2.  Disc replacement using Pro-Disc C versus fusion: a prospective randomised and controlled radiographic and clinical study.

Authors:  A Nabhan; F Ahlhelm; T Pitzen; W I Steudel; J Jung; K Shariat; O Steimer; F Bachelier; D Pape
Journal:  Eur Spine J       Date:  2006-11-14       Impact factor: 3.134

3.  Fusion versus Bryan Cervical Disc in two-level cervical disc disease: a prospective, randomised study.

Authors:  Lei Cheng; Lin Nie; Li Zhang; Yong Hou
Journal:  Int Orthop       Date:  2008-10-28       Impact factor: 3.075

4.  Surgical outcome of cervical arthroplasty using bryan(r).

Authors:  Hong-Ki Kim; Myung-Hyun Kim; Do-Sang Cho; Sung-Hak Kim
Journal:  J Korean Neurosurg Soc       Date:  2009-12-31

5.  Primary and coupled motions after cervical total disc replacement using a compressible six-degree-of-freedom prosthesis.

Authors:  A G Patwardhan; M N Tzermiadianos; P P Tsitsopoulos; L I Voronov; S M Renner; M L Reo; G Carandang; K Ritter-Lang; R M Havey
Journal:  Eur Spine J       Date:  2010-09-24       Impact factor: 3.134

6.  Is cervical disc arthroplasty superior to fusion for treatment of symptomatic cervical disc disease? A meta-analysis.

Authors:  Si Yin; Xiao Yu; Shuangli Zhou; Zhanhai Yin; Yusheng Qiu
Journal:  Clin Orthop Relat Res       Date:  2013-02-07       Impact factor: 4.176

7.  Artificial discs for lumbar and cervical degenerative disc disease -update: an evidence-based analysis.

Authors: 
Journal:  Ont Health Technol Assess Ser       Date:  2006-04-01

Review 8.  Comparison of artificial cervical arthroplasty versus anterior cervical discectomy and fusion for one-level cervical degenerative disc disease: a meta-analysis of randomized controlled trials.

Authors:  Jiaquan Luo; Sheng Huang; Min Gong; Xuejun Dai; Manman Gao; Ting Yu; Zhiyu Zhou; Xuenong Zou
Journal:  Eur J Orthop Surg Traumatol       Date:  2014-07-18

9.  Early clinical experience with the mobi-C disc prosthesis.

Authors:  Sang Hyun Kim; Hyun Chul Shin; Dong Ah Shin; Keung Nyun Kim; Do Heum Yoon
Journal:  Yonsei Med J       Date:  2007-06-30       Impact factor: 2.759

10.  Effect of cervical dynamics on adjacent segment degeneration after anterior cervical fusion with cages.

Authors:  Ahmed Elsawaf; Luciano Mastronardi; Raffaelino Roperto; Alessandro Bozzao; Manuela Caroli; Luigi Ferrante
Journal:  Neurosurg Rev       Date:  2008-10-10       Impact factor: 3.042

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.