| Literature DB >> 15636516 |
Takashi Yoshitake1, Ryuji Iizuka, Kaoru Fujino, Osamu Inoue, Kenji Yamagata, Hitoshi Nohta, Masatoshi Yamaguchi.
Abstract
An improved high-performance liquid-chromatographic (HPLC) assay for the simultaneous determination of serotonin (5-HT) and 5-hydroxyindole-3-acetic acid (5-HIAA) in human urine is reported. Following the automated precolumn derivatization of urinary 5-HT and 5-HIAA with benzylamine, the derivatives are separated by isocratic elution on a reversed-phase C18 semi-microbore column, and are fluorometrically detected at an excitation wavelength of 345 nm and an emission wavelength of 480 nm. The detection limits of 5-HT and 5-HIAA are 7 and 5 nmol/l in urine (0.7 and 0.5 fmol/20-microl injection). The proposed automatic method permits a highly selective and sensitive determination of 5-HT and 5-HIAA in human urine without any sample purification.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15636516 DOI: 10.2116/analsci.20.1687
Source DB: PubMed Journal: Anal Sci ISSN: 0910-6340 Impact factor: 2.081