Literature DB >> 15624521

Accuracy of microCT in the quantitative determination of the degree and distribution of mineralization in developing bone.

L Mulder1, J H Koolstra, T M G J Van Eijden.   

Abstract

PURPOSE: To evaluate the accuracy and applicability of a commercially available microCT system for comparative measurements of the degree and distribution of mineralization of developing bone.
MATERIAL AND METHODS: Homogeneous K2HPO4 solutions with different concentrations (range 0-800 mg/cm3) were used to assess the accuracy of a microCT system equipped with a polychromatic X-ray source. Both high (45 kV) and low (70 kV) tube peak voltages were explored. The resulting attenuation was compared with calculated theoretical attenuation values to estimate the accuracy. As an example of its applicability, the method was used to assess changes in the degree of mineralization of various regions of the mandible from two pigs of different developmental age.
RESULTS: On average, the estimated error of the measured linear attenuation was 10% or less. Accuracy was dependent on the average mineral concentration, the size of the sample, and the energy of the X-ray beam. The accuracy of the microCT system appeared sufficient to distinguish regional differences in the degree of mineralization within and between specimens of developing mandibular bone. Furthermore, the resolution of the system allowed identification of different degrees of mineralization within trabeculae.
CONCLUSION: Accuracy of microCT with polychromatic radiation can be considered adequate for assessment of the degree of mineralization of developing bone. Therefore, this method provides a three-dimensional means by which to simultaneously investigate the bone structure as well as the degree of mineralization during development in a non-destructive manner and with high resolution.

Entities:  

Mesh:

Year:  2004        PMID: 15624521     DOI: 10.1080/02841850410008171

Source DB:  PubMed          Journal:  Acta Radiol        ISSN: 0284-1851            Impact factor:   1.990


  26 in total

1.  Soft tissue variations influence HR-pQCT density measurements in a spatially dependent manner.

Authors:  Po-Hung Wu; Tanvi Gupta; Hanling Chang; Dimitry Petrenko; Anne Schafer; Galateia Kazakia
Journal:  Bone       Date:  2020-06-27       Impact factor: 4.398

Review 2.  Application of polychromatic µCT for mineral density determination.

Authors:  W Zou; N Hunter; M V Swain
Journal:  J Dent Res       Date:  2010-09-21       Impact factor: 6.116

3.  Micro-computed tomography (micro-CT) as a potential tool to assess the effect of dynamic coating routes on the formation of biomimetic apatite layers on 3D-plotted biodegradable polymeric scaffolds.

Authors:  A L Oliveira; P B Malafaya; S A Costa; R A Sousa; R L Reis
Journal:  J Mater Sci Mater Med       Date:  2007-02       Impact factor: 3.896

4.  Porosity of human mandibular condylar bone.

Authors:  G A P Renders; L Mulder; L J van Ruijven; T M G J van Eijden
Journal:  J Anat       Date:  2007-03       Impact factor: 2.610

5.  Assessment of bone tissue mineralization by conventional x-ray microcomputed tomography: comparison with synchrotron radiation microcomputed tomography and ash measurements.

Authors:  G J Kazakia; A J Burghardt; S Cheung; S Majumdar
Journal:  Med Phys       Date:  2008-07       Impact factor: 4.071

6.  Spatial mapping of mineralization with manganese-enhanced magnetic resonance imaging.

Authors:  Ingrid E Chesnick; Jose A Centeno; Todor I Todorov; Alan E Koenig; Kimberlee Potter
Journal:  Bone       Date:  2011-02-23       Impact factor: 4.398

7.  Accuracy of trabecular structure by HR-pQCT compared to gold standard μCT in the radius and tibia of patients with osteoporosis and long-term bisphosphonate therapy.

Authors:  M Krause; O Museyko; S Breer; B Wulff; C Duckstein; E Vettorazzi; C Glueer; K Püschel; K Engelke; M Amling
Journal:  Osteoporos Int       Date:  2014-02-25       Impact factor: 4.507

8.  Effect of intraspecimen spatial variation in tissue mineral density on the apparent stiffness of trabecular bone.

Authors:  Narges Kaynia; Elaine Soohoo; Tony M Keaveny; Galateia J Kazakia
Journal:  J Biomech Eng       Date:  2015-01       Impact factor: 2.097

9.  Specimen size and porosity can introduce error into microCT-based tissue mineral density measurements.

Authors:  Roberto J Fajardo; Esther Cory; Nipun D Patel; Ara Nazarian; Andres Laib; Rajaram K Manoharan; James E Schmitz; Jeremy M DeSilva; Laura M MacLatchy; Brian D Snyder; Mary L Bouxsein
Journal:  Bone       Date:  2008-09-10       Impact factor: 4.398

10.  Effects of PTH treatment on tibial bone of ovariectomized rats assessed by in vivo micro-CT.

Authors:  J E M Brouwers; B van Rietbergen; R Huiskes; K Ito
Journal:  Osteoporos Int       Date:  2009-03-05       Impact factor: 4.507

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.