Lasse Ylianttila1, Reijo Visuri, Laura Huurto, Kari Jokela.
Abstract
The suitability of a new technology single-monochromator diode array spectroradiometer for UV-radiation safety measurements, in particular for sunbed measurements, was evaluated. The linearity, cosine response, temperature response, wavelength scale, stray-light and slit function of the spectroradiometer were determined and their effects on the measurement accuracy evaluated. The main error sources were stray-light and nonideal cosine response, for which correction methods are presented. Without correction, the stray-light may reduce the accuracy of the measurement excessively, particularly in the UV-B range. The expanded uncertainty of the corrected UV measurements is estimated to be 14%, which is confirmed with the comparative measurements carried out with a well-characterized double-monochromator spectroradiometer. The measurement accuracy is sufficient for sunbed measurements, provided that all corrections described above have been done and the user of the instrument has a good understanding of the instrument's operating principles and potential error sources. If these requirements are met, the tested spectroradiometer improves and facilitates market surveillance field measurements of sunbeds.Mesh:
Year: 2005 PMID: 15623317 DOI: 10.1562/2004-06-02-RA-184
Source DB: PubMed Journal: Photochem Photobiol ISSN: 0031-8655 Impact factor: 3.421