Literature DB >> 15598462

Measurements of constraint of total knee replacement.

H Haider1, P S Walker.   

Abstract

Measurement of the constraint of total knee components in a test machine provides an objective method of describing the laxity and stability characteristics of the implant itself, independent of the knee joint into which it would be implanted. A special fixture was designed and fitted to a Bionix multi-channel loading machine. The test consisted of applying a compressive load, applying a cyclic AP force or internal-external torque, and measuring all of the displacements and rotations. Three different commonly-used TKR's showed widely different constraint characteristics. In the cyclic AP test, along with the cyclic AP displacement, displacements and rotations occurred in the other directions. This indicated that all degrees of freedom should be free to move, otherwise anomalous results would be obtained. The paper concludes with recommendations for standardized constraint tests.

Mesh:

Year:  2005        PMID: 15598462     DOI: 10.1016/j.jbiomech.2004.02.014

Source DB:  PubMed          Journal:  J Biomech        ISSN: 0021-9290            Impact factor:   2.712


  4 in total

1.  The effect of geometric variations in posterior-stabilized knee designs on motion characteristics measured in a knee loading machine.

Authors:  Peter S Walker; Michael T Lowry; Anoop Kumar
Journal:  Clin Orthop Relat Res       Date:  2014-01       Impact factor: 4.176

2.  Experimental testing of total knee replacements with UHMW-PE inserts: impact of severe wear test conditions.

Authors:  Carmen Zietz; Joern Reinders; Jens Schwiesau; Alexander Paulus; Jan Philippe Kretzer; Thomas Grupp; Sandra Utzschneider; Rainer Bader
Journal:  J Mater Sci Mater Med       Date:  2015-02-26       Impact factor: 3.896

3.  Is there a gold standard for TKA tibial component rotational alignment?

Authors:  Erin E Hutter; Jeffrey F Granger; Matthew D Beal; Robert A Siston
Journal:  Clin Orthop Relat Res       Date:  2013-02-08       Impact factor: 4.176

4.  The superficial medial collateral ligament is the primary medial restraint to knee laxity after cruciate-retaining or posterior-stabilised total knee arthroplasty: effects of implant type and partial release.

Authors:  Kiron K Athwal; Hadi El Daou; Christoph Kittl; Andrew J Davies; David J Deehan; Andrew A Amis
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2015-10-30       Impact factor: 4.342

  4 in total