[Secondary stability of cemented and non-cemented acetabular implants ex-vivo under dynamic load].

Jie Wu1, K-H Widmer, B Zurfluh, Beat Goepfert, Erxin W Morscher.   

Abstract

OBJECTIVE: To measure the bone-prosthetic implant interface micromovement during the application of physiological load by using a material testing system (MTS).
METHODS: The cadaveric hip specimens were used to simulate a single leg stance and the joint in the neutral position. Micromovement was recorded via a 3-dimensional transducer in the acetabula of postmortem specimens, which had been preserved in formalin. The study data of the cemented and uncemented prosthsis refereed to the lone-term clinical process and the radiological status and experimental results.
RESULTS: Cemented cups showed higher transverse relative motion up to 90 microm, whereas the maximum transverse movement of the non-cemented cup was 60 microm. Orthogonal motion perpendicular to the implant surface showed compression for all cups at all sites.
CONCLUSION: The results indicate that there are large differences in survival time between 2 groups. That could not be compared statistically in secondary stability. Nevertheless, according to the results, the amount of micromotion of press-fit cup is relatively less than that of cemented polyethylene cup, which is instrumental in bone ingrowth and secondary stability.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15598369

Source DB:  PubMed          Journal:  Zhonghua Wai Ke Za Zhi        ISSN: 0529-5815


  1 in total

1.  3D video-based deformation measurement of the pelvis bone under dynamic cyclic loading.

Authors:  Beat Göpfert; Zdzislaw Krol; Marie Freslier; Andreas H Krieg
Journal:  Biomed Eng Online       Date:  2011-07-17       Impact factor: 2.819

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.