Literature DB >> 15595337

Potential for Personal Digital Assistant interference with implantable cardiac devices.

Jeffrey L Tri1, Jane M Trusty, David L Hayes.   

Abstract

OBJECTIVE: To determine whether the wireless local area network (WLAN) technology, specifically the Personal Digital Assistant (PDA), interferes with implantable cardiac pacemakers and defibrillators.
MATERIAL AND METHODS: Various pacemakers and defibrillators were tested in vitro at the Mayo Clinic in Rochester, Minn, between March 6 and July 30, 2003. These cardiac devices were exposed to an HP Compaq IPAQ PDA fitted with a Cisco Aironet WLAN card. Initial testing was designed to show whether the Aironet card radiated energy in a consistent pattern from the antenna of the PDA to ensure that subsequent cardiac device testing would not be affected by the orientation of the PDA to the cardiac device. Testing involved placing individual cardiac devices in a simulator and uniformly exposing each device at its most sensitive programmable value to the WLAN card set to maximum power.
RESULTS: During testing with the Cisco WLAN Aironet card, all devices programmed to the unipolar or bipolar configuration single- or dual-chamber mode had normal pacing and sensing functions and exhibited no effects of electromagnetic interference except for 1 implantable cardioverter-defibrillator (ICD). This aberration was determined to relate to the design of the investigators' testing apparatus and not to the output of the PDA. The ICD device appropriately identified and labeled the electromagnetic aberration as "noise."
CONCLUSIONS: We documented no electromagnetic interference caused by the WLAN technology by using in vitro testing of pacemakers and ICDs; however, testing ideally should be completed in vivo to confirm the lack of any clinically important interactions.

Entities:  

Mesh:

Year:  2004        PMID: 15595337     DOI: 10.4065/79.12.1527

Source DB:  PubMed          Journal:  Mayo Clin Proc        ISSN: 0025-6196            Impact factor:   7.616


  3 in total

1.  Evaluation of electromagnetic interference and exposure assessment from s-health solutions based on Wi-Fi devices.

Authors:  Silvia de Miguel-Bilbao; Erik Aguirre; Peio Lopez Iturri; Leire Azpilicueta; José Roldán; Francisco Falcone; Victoria Ramos
Journal:  Biomed Res Int       Date:  2015-01-06       Impact factor: 3.411

Review 2.  Contraindications for video capsule endoscopy.

Authors:  Dirk Bandorski; Niehls Kurniawan; Peter Baltes; Reinhard Hoeltgen; Matthias Hecker; Dominik Stunder; Martin Keuchel
Journal:  World J Gastroenterol       Date:  2016-12-07       Impact factor: 5.742

3.  Cardiac device implantation in the United States from 1997 through 2004: a population-based analysis.

Authors:  Chunliu Zhan; William B Baine; Artyom Sedrakyan; Claudia Steiner
Journal:  J Gen Intern Med       Date:  2008-01       Impact factor: 5.128