Transcutaneous bilirubinometry with the Bilicheck in very premature newborns.

W A Willems1, L M van den Berg, H de Wit, A Molendijk.   

Abstract

OBJECTIVES: To investigate the potential advantages of use of the Bilicheck in the very preterm population, with special emphasis on the effect of possible adverse skin conditions on the accuracy of the measurements. In addition we estimated the potential for safe reduction of the number of blood samples taken for serum bilirubin determinations by introduction of the Bilicheck into the neonatal intensive-care unit (NICU).
METHODS: Total serum bilirubin (TSB) and transcutaneous bilirubin (TcB) were determined in very preterm newborns (gestational age < 30 weeks). To assess the agreement between TSB and TcB values, Bland-Altman plots were analyzed. Accuracy and (intra-device) imprecision were determined by statistical analysis. A screening model was developed to estimate the potential for safe reduction of the number of blood samples taken for TSB determination.
RESULTS: Correlations between TcB and TSB values varied between 0.86 and 0.88 and all were statistically significant. Bland-Altman plots and statistical analysis showed that the agreement between TcB and TSB measurements was largest for the group with good skin conditions, as expected. The Bilicheck device had an acceptable level of intra-device imprecision (2.29 +/- 13.51 micromol/l). Applying the screening model to our entire study population, 35 of the 93 TSB measurements (38%) could have been saved.
CONCLUSIONS: The Bilicheck is a screening device with the potential to reliably indicate hyperbilirubinemia in very preterm infants. Caution is required when skin measurements are performed in the presence of peripheral edema and/or a poor peripheral circulation. Its application in the NICU environment has the potential to reduce the number of blood samples by 40%.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15590448     DOI: 10.1080/14767050400013818

Source DB:  PubMed          Journal:  J Matern Fetal Neonatal Med        ISSN: 1476-4954


  8 in total

1.  Screening neonatal jaundice based on the sclera color of the eye using digital photography.

Authors:  Terence S Leung; Karan Kapur; Ashley Guilliam; Jade Okell; Bee Lim; Lindsay W MacDonald; Judith Meek
Journal:  Biomed Opt Express       Date:  2015-10-23       Impact factor: 3.732

2.  Interscapular site for transcutaneous bilirubin measurement in preterm infants: a better and safer screening site.

Authors:  A Yaser; L Tooke; N Rhoda
Journal:  J Perinatol       Date:  2014-01-09       Impact factor: 2.521

3.  Accuracy of transcutaneous bilirubin measurement in preterm low-birth-weight neonates.

Authors:  Deepak Chawla; Suksham Jain; Gurjit Kaur; Vikas Sinhmar; Vishal Guglani
Journal:  Eur J Pediatr       Date:  2013-08-17       Impact factor: 3.183

4.  Transcutaneous bilirubin levels in newborns <35 weeks' gestation.

Authors:  M J Maisels; M P Coffey; E Kring
Journal:  J Perinatol       Date:  2015-06-25       Impact factor: 2.521

5.  Diagnostic usefulness of transcutaneous bilirubinometry in very preterm newborns.

Authors:  Zohreh Badiee; Majid Mohammadizadeh; Masih Shamee
Journal:  Int J Prev Med       Date:  2012-04

6.  Should transcutaneous bilirubin be measured in preterm infants receiving phototherapy? The relationship between transcutaneous and total serum bilirubin in preterm infants with and without phototherapy.

Authors:  Christian V Hulzebos; Deirdre E Vader-van Imhoff; Arend F Bos; Peter H Dijk
Journal:  PLoS One       Date:  2019-06-14       Impact factor: 3.240

7.  The forehead is a better site than the sternum to check transcutaneous bilirubin during phototherapy in sick infants.

Authors:  Jaesung Jeon; Gina Lim; Ki Won Oh; Na Mi Lee; Hye Won Park; Mi Lim Chung
Journal:  BMC Pediatr       Date:  2020-12-05       Impact factor: 2.125

8.  Comparison of a new transcutaneous bilirubinometer (Bilimed) with serum bilirubin measurements in preterm and full-term infants.

Authors:  Tanja Karen; Hans Ulrich Bucher; Jean-Claude Fauchère
Journal:  BMC Pediatr       Date:  2009-11-12       Impact factor: 2.125

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.