The long-term stability of amorphous silicon flat panel imaging devices for dosimetry purposes.

R J W Louwe1, L N McDermott, J J Sonke, R Tielenburg, M Wendling, M B van Herk, B J Mijnheer.   

Abstract

This study was carried out to determine the stability of the response of amorphous silicon (a-Si)-flat panel imagers for dosimetry applications. Measurements of the imager's response under reference conditions were performed on a regular basis for four detectors of the same manufacturer. We found that the ambient temperature influenced the dark-field, while the gain of the imager signal was unaffected. Therefore, temperature fluctuations were corrected for by applying a "dynamic" darkfield correction. This correction method also removed the influence of a small, irreversible increase of the dark-field current, which was equal to 0.5% of the dynamic range of the imager per year and was probably caused by mild radiation damage to the a-Si array. By applying a dynamic dark-field correction, excellent stability of the response over the entire panel of all imagers of 0.5% (1 SD) was obtained over an observation period up to 23 months. However, two imagers had to be replaced after several months. For one imager, an image segment stopped functioning, while the image quality of the other imager degraded significantly. We conclude that the tested a-Si EPIDs have a very stable response and are therefore well suited for dosimetry. We recommend, however, applying quality assurance tests dedicated to both imaging and dosimetry.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15587651     DOI: 10.1118/1.1803751

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  19 in total

1.  A generalized calibration procedure for in vivo transit dosimetry using siemens electronic portal imaging devices.

Authors:  Andrea Fidanzio; Francesca Greco; Laura Gargiulo; Savino Cilla; Domenico Sabatino; Massimo Cappiello; Cinzia Di Felice; Elisabetta Di Castro; Luigi Azario; Mariateresa Russo; Luciano Pompei; Guido D'Onofrio; Angelo Piermattei
Journal:  Med Biol Eng Comput       Date:  2010-11-04       Impact factor: 2.602

2.  Practical guidelines for routine intensity-modulated radiotherapy verification: pre-treatment verification with portal dosimetry and treatment verification with in vivo dosimetry.

Authors:  A J Vinall; A J Williams; V E Currie; A Van Esch; D Huyskens
Journal:  Br J Radiol       Date:  2010-11       Impact factor: 3.039

3.  Assessment of a 2D electronic portal imaging devices-based dosimetry algorithm for pretreatment and in-vivo midplane dose verification.

Authors:  Ali Jomehzadeh; Parvaneh Shokrani; Mohammad Mohammadi; Alireza Amouheidari
Journal:  Adv Biomed Res       Date:  2016-11-28

4.  A method for deconvolution of integrated electronic portal images to obtain incident fluence for dose reconstruction.

Authors:  Wendel Dean Renner; Kevin Norton; Timothy Holmes
Journal:  J Appl Clin Med Phys       Date:  2005-11-21       Impact factor: 2.102

5.  Feasibility of using two-dimensional array dosimeter for in vivo dose reconstruction via transit dosimetry.

Authors:  Heeteak Chung; Jonathan Li; Sanjiv Samant
Journal:  J Appl Clin Med Phys       Date:  2011-04-08       Impact factor: 2.102

6.  Intensity-modulated radiation therapy dose verification using fluence and portal imaging device.

Authors:  Iori Sumida; Hajime Yamaguchi; Indra J Das; Hisao Kizaki; Keiko Aboshi; Mari Tsujii; Yuji Yamada; Osamu Suzuki; Yuji Seo; Fumiaki Isohashi; Kazuhiko Ogawa
Journal:  J Appl Clin Med Phys       Date:  2016-01-08       Impact factor: 2.102

7.  Evaluation of the TrueBeam machine performance check (MPC) beam constancy checks for flattened and flattening filter-free (FFF) photon beams.

Authors:  Michael P Barnes; Peter B Greer
Journal:  J Appl Clin Med Phys       Date:  2016-11-30       Impact factor: 2.102

8.  Real-time dose reconstruction for wedged photon beams: a generalized procedure.

Authors:  A Piermattei; F Greco; A Fidanzio; L Azario; A Porcelli; S Cilla; D Sabatino; A Russo; G D'Onofrio; M Russo
Journal:  J Appl Clin Med Phys       Date:  2011-11-15       Impact factor: 2.102

9.  Determination of dosimetric leaf gap using amorphous silicon electronic portal imaging device and its influence on intensity modulated radiotherapy dose delivery.

Authors:  S Timothy Peace Balasingh; I Rabi Raja Singh; K Mohamathu Rafic; S Ebenezer Suman Babu; B Paul Ravindran
Journal:  J Med Phys       Date:  2015 Jul-Sep

10.  Establishing action levels for EPID-based QA for IMRT.

Authors:  Rebecca M Howell; Iris P N Smith; Christie S Jarrio
Journal:  J Appl Clin Med Phys       Date:  2008-06-23       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.