Literature DB >> 15580627

Sudden failure of dual channel pulse generators.

François Alesch1.   

Abstract

We report a series of four sudden hardware failures in dual channel pulse generators implanted for chronic stimulation of the subthalamic nucleus in the treatment of Parkinson's disease. In all cases, a sudden and severe deterioration of the patient's neurological condition occurred with symptoms similar to those present before surgery. In all cases, destructive analysis of the generator revealed a fracture of the wire bonds connecting the battery to the hybrid (electronic) part of the pulse generator. This fracture led to repeated Power On Resets, bringing the parameter settings back to default (factory settings). As a cause, we propose that relative movements between the hybrid and the battery led to low cycle fatigue fractures due to insufficient stiffness of the device. That three of four failures occurred when the implant was in the infracostal region makes it likely that the fractures depend on the mechanical stress applied to the device. The efficacy of the therapy resumed immediately after device replacement. As a corrective measure, the manufacturer has added epoxy between both components, increasing significantly the stiffness of the device. (c) 2004 Movement Disorder Society.

Entities:  

Mesh:

Year:  2005        PMID: 15580627     DOI: 10.1002/mds.20354

Source DB:  PubMed          Journal:  Mov Disord        ISSN: 0885-3185            Impact factor:   10.338


  7 in total

Review 1.  High frequency deep brain stimulation of the subthalamic nucleus versus continuous subcutaneous apomorphine infusion therapy: a review.

Authors:  R Carron; V Fraix; C Maineri; E Seigneuret; B Piallat; P Krack; P Pollak; A L Benabid; Stéphan Chabardès
Journal:  J Neural Transm (Vienna)       Date:  2010-12-29       Impact factor: 3.575

Review 2.  Deep Brain Stimulation Emergencies: How the New Technologies Could Modify the Current Scenario.

Authors:  Giovanni Cossu; Mariachiara Sensi
Journal:  Curr Neurol Neurosci Rep       Date:  2017-07       Impact factor: 5.081

Review 3.  Disease-specific longevity of impulse generators in deep brain stimulation and review of the literature.

Authors:  Christoph van Riesen; Georg Tsironis; Doreen Gruber; Fabian Klostermann; Patricia Krause; Gerd Helge Schneider; Andreas Kupsch
Journal:  J Neural Transm (Vienna)       Date:  2016-05-19       Impact factor: 3.575

4.  Placement of the Internal Pulse Generator for Deep Brain Stimulation in the Upper Back to Prevent Fracture of the Extension Wire due to Generator Rotation: Case Report.

Authors:  Ankur Garg; Avinash L Mohan; P Charles Garell
Journal:  Parkinsons Dis       Date:  2010-02-08

5.  Clinical and programming pattern of patients with impending deep brain stimulation power failure: a retrospective chart review.

Authors:  Raja Mehanna; Kathy M Wilson; Scott E Cooper; Andre G Machado; Hubert H Fernandez
Journal:  J Clin Mov Disord       Date:  2014-11-20

6.  Management of deep brain stimulator battery failure: battery estimators, charge density, and importance of clinical symptoms.

Authors:  Kaihan Fakhar; Erin Hastings; Christopher R Butson; Kelly D Foote; Pam Zeilman; Michael S Okun
Journal:  PLoS One       Date:  2013-03-11       Impact factor: 3.240

7.  Sudden loss of the deep brain stimulation effect with high impedance without macroscopic fracture: a case report and review of the published literature.

Authors:  Hui-Jun Yang; Ji Young Yun; Young Eun Kim; Yong Hoon Lim; Han-Joon Kim; Sun Ha Paek; Beom S Jeon
Journal:  Neuropsychiatr Dis Treat       Date:  2015-07-21       Impact factor: 2.570

  7 in total