[Automatic injectors in magnetic resonance imaging and computed tomography: pilot study on hygienic aspects].

B Buerke1, A K Sonntag, R Fischbach, W Heindel, B Tombach.   

Abstract

PURPOSE: To evaluate hygienic conditions using automatic injectors in magnetic resonance imaging (MRI) and computed tomography (CT) during clinical routine.
MATERIALS AND METHODS: The surfaces of medical devices (e. g., control console) and the palms of the technical and medical staff were microbiologically analyzed by taking imprints before and after hygienic education. In addition, the injector syringes for contrast medium (CM) and saline were checked for microbiological contamination following multiple (MRI: 14 h; CT 8 h) and single use. Furthermore, the potential of retrograde contamination from the patient along the tube was analyzed.
RESULTS: A bacterial contamination with typical dermal bacteria was documented for the surfaces of the medical devices, the palms of the technical and medical staff, and the injection syringes following multiple use (MRI: 10/10 CM syringes, 6/10 saline syringes; CT: 8/10 CM syringes, 5/10 saline syringes). Correct hand disinfection in combination with single use of syringes avoided bacterial colonization. Retrograde bacterial contamination from the patient was not observed.
CONCLUSION: Regular hygienic teaching sessions for technical and medical staff in MRI and CT departments using automatic injectors should be mandatory. Furthermore, the multiple use of syringes should be avoided until investigations addressing the potential of bacterial contamination are performed.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15573296     DOI: 10.1055/s-2004-813520

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  3 in total

1.  Hygiene precautions and the transmission of infections in radiology.

Authors:  C Bibbolino; S Pittalis; V Schininà; E Busi Rizzi; V Puro
Journal:  Radiol Med       Date:  2009-01-29       Impact factor: 3.469

2.  A quasi-experimental study to determine the effects of a multifaceted educational intervention on hand hygiene compliance in a radiography unit.

Authors:  Margaret O'Donoghue; Suk-Hing Ng; Lorna Kp Suen; Maureen Boost
Journal:  Antimicrob Resist Infect Control       Date:  2016-10-19       Impact factor: 4.887

3.  Bacterial contamination of automated MRI contrast injectors in clinical routine.

Authors:  Juliane Goebel; Joerg Steinmann; Evelyn Heintschel von Heinegg; Tobias Hestermann; Kai Nassenstein
Journal:  GMS Hyg Infect Control       Date:  2019-05-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.