BACKGROUND AND
PURPOSE: Stent implantation alone might not be sufficient to produce definitive treatment of cerebral aneurysms. Therefore, extended experimental work is needed to improve results. We show the feasibility of using an in vitro anatomically shaped elastic model for flow evaluation before and after stent implantation.
METHODS: Based on human vascular casting, an anatomic elastic internal carotid artery model, including an aneurysm on the supraclinoid portion, was manufactured. The model was connected to a circulatory loop to simulate physiological flow. After visualization of the flow by using glass particles and laser sheet translumination, the digitally recorded data were transferred for computer analysis. Intra-saccular flow pattern changes and the vortex velocity reduction induced by the stent were investigated qualitatively and quantitatively.
RESULTS: The distal neck of the aneurysm behaved as a flow divider. Therefore, it was directly exposed to the hemodynamic stress. Inside the sac, a well-defined vortex formed and progressed along the wall toward the proximal neck. After stent implantation this pattern changed significantly; the vortex appeared more dispersed and its residence time increased. The velocity reduction was 32%. Velocity peak was observed close to the distal neck in both cases.
CONCLUSION: In vitro anatomic elastic models are feasible for flow evaluation with laser sheet translumination. In our model, stent implantation resulted in hemodynamic changes that might favor the exclusion of the aneurysm from the circulation and can prevent regrowth of the aneurysmal sac.

Entities:  

Mesh:

Year:  2004        PMID: 15569741      PMCID: PMC8148744     

Source DB:  PubMed          Journal:  AJNR Am J Neuroradiol        ISSN: 0195-6108            Impact factor:   3.825


  36 in total

1.  The relation of hypertension to arterial aneurysms of the brain.

Authors:  B K BLACK; S P HICKS
Journal:  U S Armed Forces Med J       Date:  1952-12

2.  Efficacy and current limitations of intravascular stents for intracranial internal carotid, vertebral, and basilar artery aneurysms.

Authors:  G Lanzino; A K Wakhloo; R D Fessler; M L Hartney; L R Guterman; L N Hopkins
Journal:  J Neurosurg       Date:  1999-10       Impact factor: 5.115

3.  Endovascular stent placement for multiple aneurysms of the extracranial internal carotid artery: technical case report.

Authors:  M Mase; T Banno; K Yamada; H Katano
Journal:  Neurosurgery       Date:  1995-10       Impact factor: 4.654

4.  Structure of blood flow through a curved vessel with an aneurysm.

Authors:  H Niimi; Y Kawano; I Sugiyama
Journal:  Biorheology       Date:  1984       Impact factor: 1.875

5.  Analysis of slipstream flow in a wide-necked basilar artery aneurysm: evaluation of potential treatment regimens.

Authors:  S G Imbesi; C W Kerber
Journal:  AJNR Am J Neuroradiol       Date:  2001-04       Impact factor: 3.825

6.  Partially clipped intracranial aneurysm obliterated with combined stent and coil implantation.

Authors:  Pedro Lylyk; José E Cohen; Angel Ferrario; Rosana Ceratto; Carlos Miranda
Journal:  J Endovasc Ther       Date:  2002-04       Impact factor: 3.487

Review 7.  Intracranial aneurysms and arterial hypertension: a review and hypothesis.

Authors:  S Inci; R F Spetzler
Journal:  Surg Neurol       Date:  2000-06

8.  Endovascular treatment of posterior circulation aneurysms by electrothrombosis using electrically detachable coils.

Authors:  G Guglielmi; F Viñuela; G Duckwiler; J Dion; P Lylyk; A Berenstein; C Strother; V Graves; V Halbach; D Nichols
Journal:  J Neurosurg       Date:  1992-10       Impact factor: 5.115

9.  Hemodynamics in rigid and distensible saccular aneurysms: a numerical study of pulsatile flow characteristics.

Authors:  M Löw; K Perktold; R Raunig
Journal:  Biorheology       Date:  1993 May-Aug       Impact factor: 1.875

10.  Combined use of stents and coils to treat experimental wide-necked carotid aneurysms: preliminary results.

Authors:  I Szikora; L R Guterman; K M Wells; L N Hopkins
Journal:  AJNR Am J Neuroradiol       Date:  1994-06       Impact factor: 3.825

View more
  18 in total

1.  Comparison of high-resolution X-ray and micro-CT for experimental evaluation of intracranial stent prototypes: quality evaluation beyond CE mark.

Authors:  Andreas Keuler; Christian Taschner; Marc Alexander Brockmann; Hanne Boll; Katharina Förster; Lisa Lutz; Annegret Herrmann-Frank; Monika Lelgemann; Martin Schumacher
Journal:  Neuroradiology       Date:  2014-01-31       Impact factor: 2.804

2.  Evaluation of an asymmetric stent patch design for a patient specific intracranial aneurysm using Computational Fluid Dynamic (CFD) calculations in the Computed Tomography (CT) derived lumen.

Authors:  Minsuok Kim; Ciprian Ionita; Rekha Tranquebar; Kenneth R Hoffmann; Dale B Taulbee; Hui Meng; Stephen Rudin
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2006

3.  Exclusion of a cavernous aneurysm by leo stent.

Authors:  J M Pumar; J A Castiñeira; F Vazquez; M Blanco; P Lylyk
Journal:  Interv Neuroradiol       Date:  2006-06-15       Impact factor: 1.610

4.  Method to quantify flow reduction in aneurysmal cavities of lateral wall aneurysms produced by stent implants used for flow diversion.

Authors:  N Fujimura; M Ohta; G Abdo; H Ylmaz; K O Lovblad; D A Rüfenacht
Journal:  Interv Neuroradiol       Date:  2006-06-15       Impact factor: 1.610

5.  Unusual Intracranial Stent Navigation through the Circle of Willis in a Patient with Recurrent Basilar Tip Aneurysm during Stent-Assisted Coiling. A Case Report.

Authors:  A S Puri; E Erdem
Journal:  Interv Neuroradiol       Date:  2009-04-15       Impact factor: 1.610

Review 6.  Silicone models as basic training and research aid in endovascular neurointervention--a single-center experience and review of the literature.

Authors:  Srinivasan Paramasivam; Gerasimos Baltsavias; Evlampia Psatha; Georgios Matis; Anton Valavanis
Journal:  Neurosurg Rev       Date:  2014-01-25       Impact factor: 3.042

7.  Computational study for the effects of coil configuration on blood flow characteristics in coil-embolized cerebral aneurysm.

Authors:  Tomohiro Otani; Satoshi Ii; Tomoyoshi Shigematsu; Toshiyuki Fujinaka; Masayuki Hirata; Tomohiko Ozaki; Shigeo Wada
Journal:  Med Biol Eng Comput       Date:  2016-07-21       Impact factor: 2.602

8.  Endovascular image-guided interventions (EIGIs).

Authors:  Stephen Rudin; Daniel R Bednarek; Kenneth R Hoffmann
Journal:  Med Phys       Date:  2008-01       Impact factor: 4.071

9.  Effect of flow diverter porosity on intraaneurysmal blood flow.

Authors:  Luca Augsburger; Mohamed Farhat; Philippe Reymond; Edouard Fonck; Zsolt Kulcsar; Nikos Stergiopulos; Daniel A Rüfenacht
Journal:  Klin Neuroradiol       Date:  2009-08-23

10.  Neurovascular modeling: small-batch manufacturing of silicone vascular replicas.

Authors:  J Y Chueh; A K Wakhloo; M J Gounis
Journal:  AJNR Am J Neuroradiol       Date:  2009-03-25       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.