Literature DB >> 15568536

Fracture of a Y-TZP ceramic femoral head. Analysis of a fault.

G Maccauro1, C Piconi, W Burger, L Pilloni, E De Santis, F Muratori, I D Learmonth.   

Abstract

We studied factors contributing to the initiation of fracture and failure of a zirconia ceramic femoral head. The materials retrieved during a revision total hip replacement were submitted to either visual, stereomicroscopic and scanning electron microscopy (SEM) or SEM and energy-dispersive X-ray analysis. X-ray diffraction was performed in order to investigate the extent of tetragonal to monoclinic phase transition. Histological examination was performed on the periprosthetic tissues. The results showed that failure was due to the propagation during clinical use of defects which may have been introduced into the material during the processing of the ceramic, rather than those intrinsic to zirconia. The literature relating to previous failures of zirconia components is reviewed.

Entities:  

Mesh:

Year:  2004        PMID: 15568536     DOI: 10.1302/0301-620x.86b8.15012

Source DB:  PubMed          Journal:  J Bone Joint Surg Br        ISSN: 0301-620X


  6 in total

1.  On the fracture of a zirconia ball head.

Authors:  C Piconi; G Maccauro; L Pilloni; W Burger; F Muratori; H G Richter
Journal:  J Mater Sci Mater Med       Date:  2006-03       Impact factor: 3.896

Review 2.  New Knowledge about Zirconium-Ceramic as a Structural Material in Fixed Prosthodontics.

Authors:  Davor Špehar; Marko Jakovac
Journal:  Acta Stomatol Croat       Date:  2015-06

3.  Primary Total Hip Arthroplasty Using Third Generation Ceramic-Ceramic Articulation: Results after a Minimum of Three-years of Follow-up.

Authors:  Dong Woo Lee; Sung Kwan Hwang
Journal:  Hip Pelvis       Date:  2014-06-30

4.  Evaluation of Mechanical and Electrical Performance of Aging Resistance ZTA Composites Reinforced with Graphene Oxide Consolidated by SPS.

Authors:  Sergey Grigoriev; Anton Smirnov; Nestor Washington Solis Pinargote; Oleg Yanushevich; Natella Kriheli; Olga Kramar; Yuri Pristinskiy; Pavel Peretyagin
Journal:  Materials (Basel)       Date:  2022-03-25       Impact factor: 3.623

Review 5.  Fracture of ceramic bearing surfaces following total hip replacement: a systematic review.

Authors:  Francesco Traina; Marcello De Fine; Alberto Di Martino; Cesare Faldini
Journal:  Biomed Res Int       Date:  2013-06-13       Impact factor: 3.411

6.  In vivo wear performance of highly cross-linked polyethylene vs. yttria stabilized zirconia and alumina stabilized zirconia at a mean seven-year follow-up.

Authors:  Masahiro Hasegawa; Akihiro Sudo
Journal:  BMC Musculoskelet Disord       Date:  2013-05-01       Impact factor: 2.362

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.