Agreement of a new whole-blood PT/INR test using capillary samples with plasma INR determinations.

Françoise Boehlen1, Guido Reber, Philippe de Moerloose.   

Abstract

PURPOSE: The objective of the study was to compare in anticoagulated patients the international normalized ratio (INR) measured with a new capillary whole-blood device, the i-STAT Portable Clinical Analyser, with conventional plasma INR obtained from the central laboratory. PATIENTS AND METHODS: Between-cartridge variability was first determined with two lyophilized controls with INR levels of 1.60 and 2.75 (n=10). Next, in 35 patients under different intensities of oral anticoagulation, capillary blood INR was measured with two i-STAT devices and was compared to central laboratory plasma INR (Innovin reagent and BCS analyser).
RESULTS: Between-cartridge coefficients of variation were 5% (95%, CI 3.4-9.1) and 3% (95%, CI 2.1-5.5) at INR levels of 1.60 and 2.75. Mean INR difference between the two i-STAT devices was 0.1, and the correlation coefficient was 0.98. Between i-STAT and central laboratory INR, the correlation coefficient was 0.95. Bias values were 0.04, 0.2, and -0.04 at INR levels of 2.0, 2.5, and 3.5, respectively.
CONCLUSION: The INR measured with the i-STAT Portable Clinical Analyser is precise and compares well with plasma INR performed in a central laboratory.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15567464     DOI: 10.1016/j.thromres.2004.06.023

Source DB:  PubMed          Journal:  Thromb Res        ISSN: 0049-3848            Impact factor:   3.944


  1 in total

1.  Evaluation of patient perceptions and outcomes related to anticoagulation point-of-care testing in ambulatory care clinics.

Authors:  Amy N Thompson; Kelly R Ragucci; Joli D Fermo; Heather P Whitley
Journal:  Pharm Pract (Granada)       Date:  2009-03-15
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.