| Literature DB >> 15562850 |
Abstract
The pump's outflow connector of the Novel Left Ventricular Assist System I (Novel LVAS I) has been redesigned to be sutured to the infrarenal abdominal aorta either as a procedure of choice (due to its simplicity) or as an alternative in the presence of an unhealthy descending thoracic aorta. The implantation of the Novel Left Ventricular Assist System II (Novel LVAS II) requires only a single passage of the pump's inflow connector through the diaphragm. Of still greater importance, the Novel LVAS II enables a considerably shorter (20- or 22-mm) Dacron vascular graft as the outflow connector to the abdominal infrarenal aorta. The electrocardiogram-synchronized Novel LVAS II possibly ensures the most effective approach to the perfusion of visceral organs and kidneys, while avoiding both excessive mechanical stress on the blood and anatomical and functional damage to the native left ventricle.Mesh:
Year: 2004 PMID: 15562850 PMCID: PMC521772
Source DB: PubMed Journal: Tex Heart Inst J ISSN: 0730-2347