| Literature DB >> 15562468 |
Qing X Yang1, Jianli Wang, Michael B Smith, Mark Meadowcroft, Xiaoyu Sun, Paul J Eslinger, Xavier Golay.
Abstract
Geometric distortion, signal-loss, and image-blurring artifacts in echo planar imaging (EPI) are caused by frequency shifts and T(2)(*) relaxation distortion of the MR signal along the k-space trajectory due to magnetic field inhomogeneities. The EPI geometric-distortion artifact associated with frequency shift can be reduced with parallel imaging techniques such as SENSE, while the signal-loss and blurring artifacts remain. The gradient-echo slice excitation profile imaging (GESEPI) method has been shown to be successful in restoring tissue T(2)(*) relaxation characteristics and is therefore effective in reducing signal-loss and image-blurring artifacts at a cost of increased acquisition time. The SENSE and GESEPI methods are complementary in artifact reduction. Combining these two techniques produces a method capable of reducing all three types of EPI artifacts while maintaining rapid acquisition time. (c) 2004 Wiley-Liss, Inc.Entities:
Mesh:
Year: 2004 PMID: 15562468 DOI: 10.1002/mrm.20303
Source DB: PubMed Journal: Magn Reson Med ISSN: 0740-3194 Impact factor: 4.668