An investigation of the 109Cd gamma-ray induced K-x-ray fluorescence (XRF) bone-lead measurement calibration procedure.

Huiling Nie1, David R Chettle, Fiona E McNeill, Joanne M O'Meara.   

Abstract

Two sets of phantoms have been used to calibrate a 109Cd y-ray induced K-XRF bone-lead measurement system. Both sets of phantoms are made of plaster of Paris, but the calibration lines are significantly different. This results in a significant difference for the derived concentrations of bone lead for the same person using these two sets of phantoms. This study shows that the different calibration lines are due to the different compositions of the phantoms, which can then be accounted for by adjusting the parameters related to the phantom composition in spectral analysis. Bone-lead concentrations for ten lead-exposed smelter workers were computed before and after analysis modification, and the results show that the bone-lead concentrations for the same person calculated from two sets of phantoms are not significantly different, only after the modifications are incorporated. Through these investigations, it was discovered that a common practice of setting the ratio of the calcium edge amplitude to the coherent scatter amplitude as a constant is only valid when all spectra are acquired at the same system resolution. When there is a change in the resolution between spectra, it has been determined that the ratio of the calcium edge amplitude to the coherent area should instead be used as the constant factor in the analysis program.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15552425     DOI: 10.1088/0031-9155/49/19/n02

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  2 in total

1.  In vivo quantification of lead in bone with a portable x-ray fluorescence system--methodology and feasibility.

Authors:  L H Nie; S Sanchez; K Newton; L Grodzins; R O Cleveland; M G Weisskopf
Journal:  Phys Med Biol       Date:  2011-01-17       Impact factor: 3.609

2.  Methodology using a portable X-ray fluorescence device for on-site and rapid evaluation of heavy-atom contamination in wounds: a model study for application to plutonium contamination.

Authors:  Hiroshi Yoshii; Kouta Yanagihara; Hitoshi Imaseki; Tsuyoshi Hamano; Hirokuni Yamanishi; Masayo Inagaki; Yasuhiro Sakai; Nobuyuki Sugiura; Osamu Kurihara; Kazuo Sakai
Journal:  PLoS One       Date:  2014-07-10       Impact factor: 3.240

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.