Curing performance of a new-generation light-emitting diode dental curing unit.

Kim M Wiggins1, Martin Hartung, Olaf Althoff, Christine Wastian, Sumita B Mitra.   

Abstract

UNLABELLED: BACKGROUND; Recent technological advances have resulted in the marketing of high-powered, or HP, battery-operated light-emitting diode, or LED, dental curing lights. The authors examine the curing efficiency and peak polymerization temperature, or Tp, of a new HP LED curing light.
METHODS: The authors studied four visible light-curing, or VLC, units: HP LED (A), first-generation LED (B), conventional halogen (C) and high-intensity halogen (D). They determined the depth of cure, or DOC; adhesion; and Tp of three types of VLC resin-based composites after exposure to each light. The exposure times for units A and D were one-half those for units B and C.
RESULTS: The power density of unit A was 1,000 milliwatts per square centimeter, which was comparable to that of unit D with turbo charge. The DOC and adhesion attained for all three resin-based composites after being light cured by unit A for a 10-second exposure time were equivalent to those after being light cured by unit D for a 10-second exposure time and to those after being light cured by units B and C for 20-second exposure times. The resin-based composites light cured by unit A attained significantly lower Tps than did those light cured by unit D at equivalent cure, or exposure, times and by unit C at twice the cure time.
CONCLUSIONS: The authors found that Unit A effectively cured the resin-based composites at one-half the cure time of units B and C and at the same time as unit D, while maintaining low Tp. CLINICAL IMPLICATIONS: The battery-operated HP LED curing light might be an effective, time-saving alternative for clinicians to use in light curing resin-based composites.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15551990     DOI: 10.14219/jada.archive.2004.0059

Source DB:  PubMed          Journal:  J Am Dent Assoc        ISSN: 0002-8177            Impact factor:   3.634


  11 in total

1.  Influence of the LED curing source and selective enamel etching on dentin bond strength of self-etch adhesives in class I composite restorations.

Authors:  Eduardo José Souza-Junior; Cíntia Tereza Pimenta Araújo; Lúcia Trazzi Prieto; Luís Alexandre Maffei Sartini Paulillo
Journal:  Lasers Med Sci       Date:  2011-11-26       Impact factor: 3.161

2.  Does a reduction of polymerization time and bonding steps affect the bond strength of brackets?

Authors:  Timea Lamper; Stefanie Steinhäuser-Andresen; Karin C Huth; Nicoleta Ilie; Ekaterini Paschos
Journal:  Clin Oral Investig       Date:  2011-03-30       Impact factor: 3.573

3.  Clinical long-term success of contemporary nano-filled resin composites in class I and II restorations cured by LED or halogen light.

Authors:  Torsten Pflaum; Stefan Kranz; Regina Montag; Arndt Güntsch; Andrea Völpel; Robin Mills; Klaus Jandt; Bernd Sigusch
Journal:  Clin Oral Investig       Date:  2017-10-28       Impact factor: 3.573

4.  Effects of residual ethanol on the rate and degree of conversion of five experimental resins.

Authors:  Milena Cadenaro; Lorenzo Breschi; Frederick A Rueggeberg; Michael Suchko; Evan Grodin; Kelli Agee; Roberto Di Lenarda; Franklin R Tay; David H Pashley
Journal:  Dent Mater       Date:  2008-12-25       Impact factor: 5.304

5.  Curing efficacy of light emitting diodes of dental curing units.

Authors:  Seyed Mostafa Mousavinasab; Ian Meyers
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2009-03-16

6.  Comparison of Depth of Cure, Hardness and Heat Generation of LED and High Intensity QTH Light Sources.

Authors:  Sayed Mostafa Mousavinasab; Ian Meyers
Journal:  Eur J Dent       Date:  2011-07

7.  Effects of light curing method and exposure time on mechanical properties of resin based dental materials.

Authors:  A Riza Alpöz; Fahinur Ertugrul; Dilsah Cogulu; Asli Topaloglu Ak; Metin Tanoglu; Elçin Kaya
Journal:  Eur J Dent       Date:  2008-01

8.  Comparison of the bonding strengths of second- and third-generation light-emitting diode light-curing units.

Authors:  Hee-Min Lee; Sang-Cheol Kim; Kyung-Hwa Kang; Na-Young Chang
Journal:  Korean J Orthod       Date:  2016-11-14       Impact factor: 1.372

Review 9.  Factors affecting polymerization of resin-based composites: A literature review.

Authors:  Maan M AlShaafi
Journal:  Saudi Dent J       Date:  2017-03-07

10.  Comparison of Bond Strength of Metal and Ceramic Brackets Bonded with Conventional and High-Power LED Light Curing Units.

Authors:  Javad Chalipa; Yasamin Farajzadeh Jalali; Fatemeh Gorjizadeh; Pedram Baghaeian; Mohammad Hashem Hoseini; Omid Mortezai
Journal:  J Dent (Tehran)       Date:  2016-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.