Literature DB >> 15551547

Admittance models for open ended coaxial probes and their place in dielectric spectroscopy.

C Gabriel1, T Y Chan, E H Grant.   

Abstract

Starting from a rigorous formulation for the admittance of an open ended coaxial probe, this paper identifies the simplifying assumptions that are introduced to obtain more practical solutions. The predictions of the models are compared to each other and to experimental data obtained on an independently calibrated experimental set-up. The advantages and limitations of a model are determined by the theoretical derivation and the numerical solution. The model identified as most suitable for the measurement of the dielectric properties of biological materials is used to determine the relative permittivity and conductivity of muscle and skull bone in the frequency range 1 MHz to 20 GHz.

Mesh:

Substances:

Year:  1994        PMID: 15551547     DOI: 10.1088/0031-9155/39/12/004

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  A novel method to decrease electric field and SAR using an external high dielectric sleeve at 3 T head MRI: numerical and experimental results.

Authors:  Bu S Park; Sunder S Rajan; Joshua W Guag; Leonardo M Angelone
Journal:  IEEE Trans Biomed Eng       Date:  2014-10-22       Impact factor: 4.538

2.  Expanded modeling of temperature-dependent dielectric properties for microwave thermal ablation.

Authors:  Zhen Ji; Christopher L Brace
Journal:  Phys Med Biol       Date:  2011-07-26       Impact factor: 3.609

3.  A Bioimpedance-Based Device to Assess the Volume Conduction Properties of the Tongue in Neurological Disorders Affecting Bulbar function.

Authors:  Xuesong Luo; Hilda Victoria Gutierrez Pulido; Seward Rutkove; Benjamin Sanchez
Journal:  IEEE Open J Eng Med Biol       Date:  2021-10-06

4.  Open-Ended Coaxial Dielectric Probe Effective Penetration Depth Determination.

Authors:  Paul M Meaney; Andrew P Gregory; Jan Seppälä; Tapani Lahtinen
Journal:  IEEE Trans Microw Theory Tech       Date:  2016-01-29       Impact factor: 3.599

5.  Millimeter-wave emissivity as a metric for the non-contact diagnosis of human skin conditions.

Authors:  Amani Yousef Owda; Neil Salmon; Stuart William Harmer; Sergiy Shylo; Nicholas John Bowring; Nacer Ddine Rezgui; Mamta Shah
Journal:  Bioelectromagnetics       Date:  2017-08-24       Impact factor: 2.010

6.  Numerical Sensitivity Analysis for Dielectric Characterization of Biological Samples by Open-Ended Probe Technique.

Authors:  Marta Cavagnaro; Giuseppe Ruvio
Journal:  Sensors (Basel)       Date:  2020-07-04       Impact factor: 3.576

Review 7.  Open-Ended Coaxial Probe Technique for Dielectric Measurement of Biological Tissues: Challenges and Common Practices.

Authors:  Alessandra La Gioia; Emily Porter; Ilja Merunka; Atif Shahzad; Saqib Salahuddin; Marggie Jones; Martin O'Halloran
Journal:  Diagnostics (Basel)       Date:  2018-06-05

8.  Classification of rat mammary carcinoma with large scale in vivo microwave measurements.

Authors:  Emre Onemli; Sulayman Joof; Cemanur Aydinalp; Nural Pastacı Özsobacı; Fatma Ateş Alkan; Nuray Kepil; Islem Rekik; Ibrahim Akduman; Tuba Yilmaz
Journal:  Sci Rep       Date:  2022-01-10       Impact factor: 4.379

9.  Open-Ended Transmission Coaxial Probes for Sarcopenia Assessment.

Authors:  Paul Meaney; Shireen D Geimer; Roberta M diFlorio-Alexander; Robin Augustine; Timothy Raynolds
Journal:  Sensors (Basel)       Date:  2022-01-19       Impact factor: 3.576

10.  Microwave open-ended coaxial dielectric probe: interpretation of the sensing volume re-visited.

Authors:  Paul M Meaney; Andrew P Gregory; Neil R Epstein; Keith D Paulsen
Journal:  BMC Med Phys       Date:  2014-06-17
  10 in total