Corneal and conjunctival sensitivity to air stimuli.

F Stapleton1, M E Tan, E B Papas, K Ehrmann, B Golebiowski, J Vega, B A Holden.   

Abstract

AIMS: To determine the repeatability of ocular surface sensitivity to mechanical stimulation using air stimuli and the effect of contact lens (CL) wear on sensitivity.
METHODS: Repeatability: 14 subjects (24-39 years) participated. Mechanical sensitivity to warmed (34 degrees C) and ambient (20 degrees C) air was measured for the central cornea (CC), inferior cornea (IC), and inferior conjunctiva (ICON). Measurements were taken on 12 days; six morning and six afternoon measurements. Differences between sites, time of day, and stimulus temperature were evaluated. CL wear: 10 subjects (22-30 years) participated. Measurements were taken at the same time of day, either following no wear, wear of a CL of oxygen permeability [Dk] of 28 x 10(-9) [cm/s][ml O(2)/ml mm Hg] or wear of a CL of Dk 140 x 10(-9) [cm/s][ml O(2)/ml mm Hg]. Differences between sites and wear conditions were evaluated.
RESULTS: Repeatability: Sensitivity varied between sites (p<0.01), time of day (p<0.05), and stimulus temperatures (p<0.01). There were no significant differences between days. Mean thresholds for eye temperature stimuli were; CC 64.4 (SD 28.6) ml/min; IC 84.6 (40.0) ml/min; ICON 120.6 (40.4) ml/min and for ambient temperature stimuli were CC 53.9 (16.0) ml/min, IC 59.0 (20.0) ml/min; ICON 72.6 (43.7) ml/min. CL wear: Sensitivity varied between sites and wear conditions (p<0.05). Conjunctival sensitivity was increased after wear of highly oxygen permeable CLs but unaffected by wear of low oxygen permeable CLs.
CONCLUSIONS: The prototype gas aesthesiometer is able to repeatably measure ocular surface sensitivity and measurements are consistent with previously reported techniques.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15548810      PMCID: PMC1772443          DOI: 10.1136/bjo.2004.044024

Source DB:  PubMed          Journal:  Br J Ophthalmol        ISSN: 0007-1161            Impact factor:   4.638


  28 in total

1.  Reliability of the non-contact corneal aesthesiometer and its comparison with the Cochet-Bonnet aesthesiometer.

Authors:  P J Murphy; J G Lawrenson; S Patel; J Marshall
Journal:  Ophthalmic Physiol Opt       Date:  1998-11       Impact factor: 3.117

2.  Corneal surface temperature change as the mode of stimulation of the non-contact corneal aesthesiometer.

Authors:  P J Murphy; P B Morgan; S Patel; J Marshall
Journal:  Cornea       Date:  1999-05       Impact factor: 2.651

Review 3.  Esthesiometry as an indicator of corneal health.

Authors:  N A Brennan; A S Bruce
Journal:  Optom Vis Sci       Date:  1991-09       Impact factor: 1.973

4.  Measurement of corneal sensitivity to mechanical and chemical stimulation with a CO2 esthesiometer.

Authors:  C Belmonte; M C Acosta; M Schmelz; J Gallar
Journal:  Invest Ophthalmol Vis Sci       Date:  1999-02       Impact factor: 4.799

5.  A new non-contact corneal aesthesiometer (NCCA).

Authors:  P J Murphy; S Patel; J Marshall
Journal:  Ophthalmic Physiol Opt       Date:  1996-03       Impact factor: 3.117

6.  Comparison of corneal sensation following photorefractive keratectomy and laser in situ keratomileusis.

Authors:  A J Kanellopoulos; I G Pallikaris; E D Donnenfeld; S Detorakis; K Koufala; H D Perry
Journal:  J Cataract Refract Surg       Date:  1997 Jan-Feb       Impact factor: 3.351

7.  Response of sensory units with unmyelinated fibres to mechanical, thermal and chemical stimulation of the cat's cornea.

Authors:  J Gallar; M A Pozo; R P Tuckett; C Belmonte
Journal:  J Physiol       Date:  1993-08       Impact factor: 5.182

8.  Excitation by irritant chemical substances of sensory afferent units in the cat's cornea.

Authors:  C Belmonte; J Gallar; M A Pozo; I Rebollo
Journal:  J Physiol       Date:  1991-06       Impact factor: 5.182

9.  Ultrastructural organization of human corneal nerves.

Authors:  L J Müller; L Pels; G F Vrensen
Journal:  Invest Ophthalmol Vis Sci       Date:  1996-03       Impact factor: 4.799

10.  Variations in corneal sensitivity with hydrogel contact lenses.

Authors:  M J Velasco; F J Bermúdez; J Romero; E Hita
Journal:  Acta Ophthalmol (Copenh)       Date:  1994-02
View more
  12 in total

1.  Corneal and conjunctival sensitivity in intolerant contact lens wearers.

Authors:  Fiona Stapleton; Blanka Golebiowski; Cheryl Skotnitsky; Maxine E Tan; Brien A Holden
Journal:  J Optom       Date:  2014-06-07

2.  The TFOS International Workshop on Contact Lens Discomfort: report of the subcommittee on neurobiology.

Authors:  Fiona Stapleton; Carl Marfurt; Blanka Golebiowski; Mark Rosenblatt; David Bereiter; Carolyn Begley; Darlene Dartt; Juana Gallar; Carlos Belmonte; Pedram Hamrah; Mark Willcox
Journal:  Invest Ophthalmol Vis Sci       Date:  2013-10-18       Impact factor: 4.799

Review 3.  TFOS DEWS II pain and sensation report.

Authors:  Carlos Belmonte; Jason J Nichols; Stephanie M Cox; James A Brock; Carolyn G Begley; David A Bereiter; Darlene A Dartt; Anat Galor; Pedram Hamrah; Jason J Ivanusic; Deborah S Jacobs; Nancy A McNamara; Mark I Rosenblatt; Fiona Stapleton; James S Wolffsohn
Journal:  Ocul Surf       Date:  2017-07-20       Impact factor: 5.033

4.  Degree of corneal anesthesia after topical application of 0.4% oxybuprocaine ophthalmic solution in normal equids.

Authors:  Erika Little; Kathy Yvorchuk-St Jean; William Little; Fortune Sithole; Guy St Jean
Journal:  Can J Vet Res       Date:  2016-10       Impact factor: 1.310

5.  The conjunctival sensitivity in soft contact lens wearers.

Authors:  Tsutomu Igarashi; Masafumi Ono; Chiaki Fujimoto; Hisaharu Suzuki; Hiroshi Takahashi
Journal:  Int Ophthalmol       Date:  2014-09-07       Impact factor: 2.031

6.  Relationship between corneal sensitivity, corneal thickness, corneal diameter, and intraocular pressure in normal cats and cats with congenital glaucoma.

Authors:  Mary R Telle; Nickolas Chen; Daniel Shinsako; Julie A Kiland; Kazuya Oikawa; Ralph Møller Trane; Gillian J McLellan
Journal:  Vet Ophthalmol       Date:  2018-03-08       Impact factor: 1.644

7.  How Should Corneal Nerves be Incorporated Into the Diagnosis and Management of Dry Eye?

Authors:  Sneh Patel; Divy Mehra; Kimberly Cabrera; Anat Galor
Journal:  Curr Ophthalmol Rep       Date:  2021-05-20

8.  Hypersensitivity to Cold Stimuli in Symptomatic Contact Lens Wearers.

Authors:  Ping Situ; Trefford Simpson; Carolyn Begley
Journal:  Optom Vis Sci       Date:  2016-08       Impact factor: 1.973

9.  A Method to Calibrate the Carbon Dioxide (Chemical) Stimuli of Pneumatic Esthesiometer Externally.

Authors:  Varadharajan Jayakumar; Trefford L Simpson
Journal:  Transl Vis Sci Technol       Date:  2019-09-11       Impact factor: 3.283

Review 10.  Corneal sensitivity in keratoconus: a review of the literature.

Authors:  Leopoldo Spadea; Serena Salvatore; Enzo Maria Vingolo
Journal:  ScientificWorldJournal       Date:  2013-10-31
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.