Literature DB >> 15544710

Screening for fever by remote-sensing infrared thermographic camera.

Lung-Sang Chan1, Giselle T Y Cheung, Ian J Lauder, Cyrus R Kumana, Ian J Lauder.   

Abstract

BACKGROUND: Following the severe acute respiratory syndrome (SARS) outbreak, remote-sensing infrared thermography (IRT) has been advocated as a possible means of screening for fever in travelers at airports and border crossings, but its applicability has not been established. We therefore set out to evaluate (1) the feasibility of IRT imaging to identify subjects with fever, and (2) the optimal instrumental configuration and validity for such testing.
METHODS: Over a 20-day inclusive period, 176 subjects (49 hospital inpatients without SARS or suspected SARS, 99 health clinic attendees and 28 healthy volunteers) were recruited. Remotely sensed IRT readings were obtained from various parts of the front and side of the face (at distances of 1.5 and 0.5 m), and compared to concurrently determined body temperature measurements using conventional means (aural tympanic IRT and oral mercury thermometry). The resulting data were submitted to linear regression/correlation and sensitivity analyses. All recruits gave prior informed consent and our Faculty Institutional Review Board approved the protocol.
RESULTS: Optimal correlations were found between conventionally measured body temperatures and IRT readings from (1) the front of the face at 1.5m with the mouth open (r=0.80), (2) the ear at 0.5 m (r=0.79), and (3) the side of the face at 1.5m (r=0.76). Average IRT readings from the forehead and elsewhere were 1 degrees C to 2 degrees C lower and correlated less well. Ear IRT readings at 0.5 m yielded the narrowest confidence intervals and could be used to predict conventional body temperature readings of < or = 38 degrees C with a sensitivity and specificity of 83% and 88% respectively.
CONCLUSIONS: IRT readings from the side of the face, especially from the ear at 0.5 m, yielded the most reliable, precise and consistent estimates of conventionally determined body temperatures. Our results have important implications for walk-through IRT scanning/screening systems at airports and border crossings, particularly as the point prevalence of fever in such subjects would be very low.

Entities:  

Mesh:

Year:  2004        PMID: 15544710     DOI: 10.2310/7060.2004.19102

Source DB:  PubMed          Journal:  J Travel Med        ISSN: 1195-1982            Impact factor:   8.490


  19 in total

1.  Effect of wearing an N95 filtering facepiece respirator on superomedial orbital infrared indirect brain temperature measurements.

Authors:  Travis DiLeo; Raymond J Roberge; Jung-Hyun Kim
Journal:  J Clin Monit Comput       Date:  2016-01-13       Impact factor: 2.502

2.  Early assessment of the efficacy of digital infrared thermal imaging in pediatric extremity trauma.

Authors:  Cicero T Silva; Nausheen Naveed; Syed Bokhari; Kenneth E Baker; Lawrence H Staib; Saad M Ibrahim; Karl Muchantef; Thomas R Goodman
Journal:  Emerg Radiol       Date:  2012-02-24

3.  Challenges to Global Implementation of Infrared Thermography Technology: Current Perspective.

Authors:  Michael Shterenshis
Journal:  Cent Asian J Glob Health       Date:  2017-10-30

4.  FluNet: An AI-Enabled Influenza-Like Warning System.

Authors:  Ryan J Ward; Fred Paul Mark Jjunju; Isa Kabenge; Rhoda Wanyenze; Elias J Griffith; Noble Banadda; Stephen Taylor; Alan Marshall
Journal:  IEEE Sens J       Date:  2021-09-16       Impact factor: 3.301

5.  Thermal image scanning for influenza border screening: results of an airport screening study.

Authors:  Patricia C Priest; Alasdair R Duncan; Lance C Jennings; Michael G Baker
Journal:  PLoS One       Date:  2011-01-05       Impact factor: 3.240

6.  Fever screening during the influenza (H1N1-2009) pandemic at Narita International Airport, Japan.

Authors:  Hiroshi Nishiura; Kazuko Kamiya
Journal:  BMC Infect Dis       Date:  2011-05-03       Impact factor: 3.090

7.  Validity of inner canthus temperature recorded by infrared thermography as a non-invasive surrogate measure for core temperature at rest, during exercise and recovery.

Authors:  Alex Andrade Fernandes; Danilo Gomes Moreira; Ciro José Brito; Cristiano Diniz da Silva; Manuel Sillero-Quintana; Eduardo Mendonça Pimenta; Aaron J E Bach; Emerson Silami Garcia; João Carlos Bouzas Marins
Journal:  J Therm Biol       Date:  2016-09-29       Impact factor: 2.902

8.  Comparison of 3 infrared thermal detection systems and self-report for mass fever screening.

Authors:  An V Nguyen; Nicole J Cohen; Harvey Lipman; Clive M Brown; Noelle Angelique Molinari; William L Jackson; Hannah Kirking; Paige Szymanowski; Todd W Wilson; Bisan A Salhi; Rebecca R Roberts; David W Stryker; Daniel B Fishbein
Journal:  Emerg Infect Dis       Date:  2010-11       Impact factor: 6.883

9.  Infrared thermography in paediatrics: a narrative review of clinical use.

Authors:  Ruaridh Owen; Shammi Ramlakhan
Journal:  BMJ Paediatr Open       Date:  2017-09-18

10.  Clinical evaluation of fever-screening thermography: impact of consensus guidelines and facial measurement location.

Authors:  Yangling Zhou; Pejman Ghassemi; Michelle Chen; David McBride; Jon P Casamento; T Joshua Pfefer; Quanzeng Wang
Journal:  J Biomed Opt       Date:  2020-09       Impact factor: 3.170

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.