Literature DB >> 15543783

An algorithm for independent verification of Gamma Knife treatment plans.

James Beck1, Anita Berndt.   

Abstract

A formalism for independent treatment verification has been developed for Gamma Knife radiosurgery in analogy to the second checks being performed routinely in the field of external beam radiotherapy. A verification algorithm is presented, and evaluated based on its agreement with treatment planning calculations for the first 40 Canadian Gamma Knife patients. The algorithm is used to calculate the irradiation time for each shot, and the value of the dose at the maximum dose point in each calculation matrix. Data entry consists of information included on the plan printout, and can be streamlined by using an optional plan import feature. Calculated shot times differed from those generated by the treatment planning software by an average of 0.3%, with a standard deviation of 1.4%. The agreement of dose maxima was comparable with an average of -0.2% and a standard deviation of 1.3%. Consistently accurate comparisons were observed for centrally located lesions treated with a small number of shots. Large discrepancies were almost all associated with dose plans utilizing a large number of collimator plugs, for which the simplifying approximations used by the program are known to break down.

Entities:  

Mesh:

Year:  2004        PMID: 15543783     DOI: 10.1118/1.1786532

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  4 in total

1.  The variable ellipsoid modeling technique as a verification method for the treatment planning system of gamma knife radiosurgery.

Authors:  Beong Ik Hur; Byung Kwan Choi; Soon Ki Sung; Won Ho Cho; Seung Heon Cha; Chang Hwa Choi
Journal:  J Korean Neurosurg Soc       Date:  2010-02-28

2.  Validation of Gamma Knife Perfexion Dose Profile Distribution by a Modified Variable Ellipsoid Modeling Technique.

Authors:  Beong Ik Hur; Seong Jin Jin; Gyeong Rip Kim; Jong Hyeok Kwak; Young Ha Kim; Sang Weon Lee; Soon Ki Sung
Journal:  J Korean Neurosurg Soc       Date:  2021-01-01

3.  Two-year experience with the commercial Gamma Knife Check software.

Authors:  Andy Yuanguang Xu; Jagdish Bhatnagar; Greg Bednarz; Josef Novotny; John Flickinger; L Dade Lunsford; M Saiful Huq
Journal:  J Appl Clin Med Phys       Date:  2016-07-08       Impact factor: 2.102

4.  Development and validation of a comprehensive patient-specific quality assurance program for a novel stereotactic radiation delivery system for breast lesions.

Authors:  Stewart J Becker; Ying Niu; Yildirim Mutaf; Shifeng Chen; Yannick Poirier; Elizabeth M Nichols; ByongYong Yi
Journal:  J Appl Clin Med Phys       Date:  2019-12       Impact factor: 2.102

  4 in total