Evaluation of cut quality using the Amadeus microkeratome with different settings.

Mana Tehrani1, Stefanie Schäfer, H Burkhard Dick.   

Abstract

PURPOSE: To evaluate the quality of keratectomy specimens created with the Amadeus microkeratome (AMO) using scanning electron microscopy (SEM).
SETTING: Johannes Gutenberg-University Hospital, Department of Ophthalmology, Mainz, and the Department of Anatomy, Justus Liebig-University, Giessen, Germany.
METHODS: Corneal cuts were performed in freshly enucleated pig eyes using the Amadeus microkeratome with 9 combinations of oscillation rate and head-advance speed. For the cutting trials, oscillation rates of 8000, 13,000, and 18,000 rpm and head-advance speeds of 1.0, 2.5, and 4.0 mm/s were chosen. Fifty-four eyes were included, resulting in 9 groups of 6 eyes for each configuration. The surfaces and edges of the flaps were examined using SEM.
RESULTS: At fixed oscillation rates, an increase in head-advance speed led to lower quality cuts, higher surface roughness, and irregular cut edges. At fixed head-advance speeds, an increase in oscillation rates improved the cut quality, resulting in smoother surface characteristics.
CONCLUSIONS: Smooth and regular surfaces and cut edges can be achieved by choosing high oscillation rates and low head-advance speeds. Microkeratomes that enable the user to adjust the oscillation rate and head-advance speed can be used to maximize surface smoothness and improve customized refractive surgery.

Entities:  

Mesh:

Year:  2004        PMID: 15519098     DOI: 10.1016/j.jcrs.2004.03.036

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  [Femtosecond laser in refractive surgery].

Authors:  H B Dick
Journal:  Ophthalmologe       Date:  2006-12       Impact factor: 1.059

2.  "OSMO-UT-DSAEK" using THIN-C medium.

Authors:  F Bucher; S Roters; A Mellein; D Hos; L M Heindl; C Cursiefen; M Hermann
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2013-08-02       Impact factor: 3.117

3.  Influence of microkeratome parameters on the stromal bed and flap edge quality in laser in situ keratomileusis.

Authors:  Jens Heichel; Frank Wilhelm; Kathleen S Kunert; Rabea Schlueter; Ute Stuhltraeger; Thomas Hammer
Journal:  Clin Ophthalmol       Date:  2013-12-17
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.