Literature DB >> 15519097

Removal times for a dispersive and a cohesive ophthalmic viscosurgical device correlated with intraocular lens material.

Gerd U Auffarth1, Mike P Holzer, Nithi Visessook, David J Apple, Hans E Völcker.   

Abstract

PURPOSE: To investigate the removal times of ophthalmic viscosurgical devices (OVDs) with different intraocular lens (IOL) designs and materials.
SETTING: Center for Research on Ocular Therapeutics and Biodevices, Department of Ophthalmology, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA, and Heidelberg IOL & Refractive Surgery Research Group, Department of Ophthalmology, Ruprecht-Karls-University of Heidelberg, Heidelberg, Germany.
METHODS: In a standardized laboratory setup, the Miyake-Apple posterior view video technique was used to evaluate OVD removal from capsular bags in human autopsy eyes implanted with poly(methyl methacrylate) (PMMA), silicone, and acrylic IOLs. The cohesive OVD ProVisc (sodium hyaluronate 1.0%) and the dispersive OVD Viscoat (sodium hyaluronate 3.0% and chondroitin sulfate 4.0%) were stained with fluorescein for better visualization. The open-sky preparation and an Alcon Series 20000 Legacy phaco machine with a flow rate of 25 mL/min and a vacuum setting of +500 mm Hg (maximum irrigation/aspiration) were used. The time needed for complete removal of the cohesive and dispersive OVDs with each IOL type was measured and analyzed statistically.
RESULTS: The mean removal times for both OVDs were as follows: Alcon MZ60BD PMMA IOL-25.0 seconds +/- 3.7 (SD) (Viscoat), 15.9 +/- 6.9 seconds (ProVisc); Alcon AcrySof MA60BM IOL-35.5 +/- 10.0 seconds (Viscoat), 25.6 +/- 4.7 seconds (ProVisc); Chiron/Bausch & Lomb C1043 silicone IOL-46.5 +/- 10.5 seconds (Viscoat), 17.3 +/- 2.1 seconds (ProVisc); AMO SI-30 silicone IOL-33.5 +/- 3.1 seconds (Viscoat), 15.3 +/- 6.3 seconds (ProVisc); and Pharmacia 912 silicone IOL-18.3 +/- 5.8 seconds (Viscoat), 19.8 +/- 4.3 seconds (ProVisc).
CONCLUSIONS: Differences in OVD removal times were detected. The removal time for the cohesive OVD correlated with the IOL material. Overall, the time needed for complete removal was significantly longer for the dispersive OVD than for the cohesive OVD.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15519097     DOI: 10.1016/j.jcrs.2004.03.039

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  9 in total

1.  Retention and removal of a new viscous dispersive ophthalmic viscosurgical device during cataract surgery in animal eyes.

Authors:  T Oshika; F Okamoto; Y Kaji; T Hiraoka; T Kiuchi; M Sato; K Kawana
Journal:  Br J Ophthalmol       Date:  2006-04       Impact factor: 4.638

2.  Intraocular lens implantation without the use of ophthalmic viscosurgical device.

Authors:  Ugur Unsal; Gonen Baser; Mehmet Soyler
Journal:  Int Ophthalmol       Date:  2016-03-14       Impact factor: 2.031

3.  Viscoless microincision cataract surgery.

Authors:  Guy Sallet
Journal:  Clin Ophthalmol       Date:  2008-12

4.  Use of viscoelastic substance in ophthalmic surgery - focus on sodium hyaluronate.

Authors:  Tomomi Higashide; Kazuhisa Sugiyama
Journal:  Clin Ophthalmol       Date:  2008-03

5.  Dispersive viscosurgical devices demonstrate greater efficacy in protecting corneal endothelium in vitro.

Authors:  Timur Mert Yildirim; Gerd U Auffarth; Hyeck-Soo Son; Ramin Khoramnia; Donald John Munro; Patrick R Merz
Journal:  BMJ Open Ophthalmol       Date:  2019-02-16

Review 6.  Ophthalmic Viscosurgical Devices (OVDs) in Challenging Cases: a Review.

Authors:  Andreas F Borkenstein; Eva-Maria Borkenstein; Boris Malyugin
Journal:  Ophthalmol Ther       Date:  2021-10-06

7.  Ophthalmic viscosurgical device interaction with two hydrophobic acrylic intraocular lenses of different equilibrium water content.

Authors:  Gerd U Auffarth; Sonja K Schickhardt; Hui Fang; Qiang Wang; Ramin Khoramnia; Timur M Yildirim
Journal:  Sci Rep       Date:  2022-08-26       Impact factor: 4.996

8.  A case of cataract surgery without pupillary device in the eye with iridoschisis.

Authors:  Eun Ji Lee; Jin Hak Lee; Joon Young Hyon; Mee Kum Kim; Won Ryang Wee
Journal:  Korean J Ophthalmol       Date:  2008-03

9.  Changes in corneal endothelium cell characteristics after cataract surgery with and without use of viscoelastic substances during intraocular lens implantation.

Authors:  Stephan D Schulze; Thomas Bertelmann; Irena Manojlovic; Stefan Bodanowitz; Sebastian Irle; Walter Sekundo
Journal:  Clin Ophthalmol       Date:  2015-11-06
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.