Left ventricular ejection and filling rate measurement based on the automatic edge detection method of ECG-gated blood pool single-photon emission tomography.

Takahiro Higuchi1, Junichi Taki, Kenichi Nakajima, Seigo Kinuya, Masatoshi Ikeda, Masanobu Namura, Norihisa Tonami.   

Abstract

UNLABELLED: The objective of the present investigation was to determine the feasibility of assessing left ventricular systolic ejection and diastolic filling via the automatic edge detection method employing ECG-gated blood pool single-photon emission tomography (SPET GBP) data.
METHODS: Thirty-five patients, who had undergone both SPET GBP and ECG-gated equilibrium blood pool scintigraphy by the planar method (planar GBP), were enrolled in this study. Planar GBP was performed with a single-headed gamma camera. From the left anterior oblique projection, data were acquired at 24 frames/cardiac cycle with ECG-gating during the equilibrium state. SPET GBP was conducted utilizing a triple-headed gamma camera, with 60 projection views over 360 degrees by 60 sec per view, in 16 frames/cardiac cycle. In each frame, left ventricular volume was determined by automatic edge detection employing a quantitative gated SPET program. Additionally, the time-volume curve was fitted by the 4th harmonics of Fourier transform. Ejection fraction (EF, %), peak ejection rate (PER, /sec), peak filling rate (PFR, /sec) and mean filling rate during the initial one-third of diastolic time (1/3 FRm) were calculated from the fitted curve. These parameters were also calculated with planar GBP data. RESULT: Left ventricular ejection and filling parameters were calculated by SPET GBP with the automatic edge detection program for all patient data. Correlation coefficients of EF, PER, PFR and 1/3 FRm between SPET and planar GBP were 0.91 (p < 0.001), 0.82 (p < 0.001), 0.78 (p < 0.001) and 0.74 (p < 0.001), respectively.
CONCLUSION: Ejection and filling rates can be calculated using SPET GBP with the edge-detection software. These parameters displayed significant correlations with those values obtained via planar GBP. Additional studies are warranted to determine the reliability of parameters with SPET GBP.

Entities:  

Mesh:

Year:  2004        PMID: 15515751     DOI: 10.1007/bf02984568

Source DB:  PubMed          Journal:  Ann Nucl Med        ISSN: 0914-7187            Impact factor:   2.668


  3 in total

Review 1.  SPECT radionuclide angiography: it is time for a consensus statement.

Authors:  Doumit Daou
Journal:  Eur J Nucl Med Mol Imaging       Date:  2007-11       Impact factor: 9.236

2.  Development of a new Python-based cardiac phantom for myocardial SPECT imaging.

Authors:  Osama S Hanafy; Magdy M Khalil; Ibrahim M Khater; Haitham S Mohammed
Journal:  Ann Nucl Med       Date:  2020-10-17       Impact factor: 2.668

3.  Depressed Exercise Peak Ejection Rate Detected on Ambulatory Radionuclide Monitoring Reflects End-Stage Cardiac Inotropic Reserve and Predicts Mortality in Ischaemic Cardiomyopathy.

Authors:  Gian Piero Carboni
Journal:  Cardiol Res       Date:  2012-07-20
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.