Peleg Ben-Galim1, Ely L Steinberg, Yishai Rosenblatt, Nata Parnes, Aharon Menahem, Ron Arbel.
Abstract
The WizAir-DVT is a miniature, lightweight (690 g), battery-operated and mobile intermittent pneumatic compression device (ICD), which enables continuous intraoperative use and immediate patient mobilization postoperatively. We compared its efficacy with a commonly used ICD, the Kendall SCD. Peak femoral vein flow velocity was measured in 20 apparently healthy volunteers at rest and with each device: we found no significant differences between them. A second prospective, randomized, clinical trial was used to compare the efficiency of the device in preventing deep venous thrombosis (DVT) after joint replacement in 50 patients (n=25/group). None developed DVT. Doppler ultrasonography revealed no significant differences. The WizAir-DVT antithrombotic compression device is as safe and effective as the Kendall SCD.Entities:
Mesh:
Year: 2004 PMID: 15513491 DOI: 10.1080/00016470410001466