Abstract

AIM: The authors assess a modified Greenfield filter (GF) for the long-term patency, filter tilting and histopathologic alterations of the inferior vena cava (IVC).
METHODS: Adult sheep (n=7) underwent modified GF placement in the IVC. Cavograms were obtained every 3 months and pulmonary angiography at 12 months. Histopathologic and scanning electron microscopy (SEM) analyses were performed on the IVC explanted at 12 months.
RESULTS: Cavograms showed that all IVC were patent at the end of the study. Filter tilting occurred in 2/7 animals and extrusion of struts was not observed. Macroscopic examination at explantation showed minimal venous wall thickening. Microscopic examination showed minimal IVC fibrosis and intimal hyperplasia. SEM showed endothelium on the IVC surface at the filter implantation site and a presumed endothelial layer covering partially or totally the struts. The interface filter-IVC was covered by deposits of leucocytes and platelets. No signs of pulmonary embolism were found in all pulmonary angiograms of both groups.
CONCLUSION: The modified filter presented good biocompatibility, stability and absence of thrombogenicity at 12 months. It presented low tendency to tilting and extrusion of struts. The long-term histopathologic alterations in vena caval wall were minimal and the appearance of the studied filters in the IVC was similar to stents placed in the arterial system.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15507896

Source DB:  PubMed          Journal:  Int Angiol        ISSN: 0392-9590            Impact factor:   2.789


  2 in total

Review 1.  Endoluminal occlusion devices: technology update.

Authors:  Tobias Zander; Samantha Medina; Guillermo Montes; Lourdes Nuñez-Atahualpa; Michel Valdes; Manuel Maynar
Journal:  Med Devices (Auckl)       Date:  2014-12-01

2.  In vivo evaluation of a new vena cava filter.

Authors:  Gilberto do Nascimento Galego; Pierre Galvagni Silveira; Júlia Jochen Broering; Eduardo da Silva Eli; Marcelo Peixer Corbellini; Amir Antônio Martins de Oliveira
Journal:  J Vasc Bras       Date:  2016 Apr-Jun
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.