Radiological parameters and radiation doses of patients undergoing abdomen, pelvis and lumbar spine X-ray examinations in three Nigerian hospitals.

F O Ogundare1, C Z Uche, F A Balogun.   

Abstract

Thermoluminescent dosemeters (TLDs) have been used to measure the entrance surface doses (ESDs) of patients undergoing pelvis, abdomen and lumbar spine diagnostic X-ray examinations in Nigeria. A total of three public hospitals and 171 patients were included in this investigation. The ages of the patients involved were from 40 years to 85 years, while their weights ranged from 64 kg to 73 kg. Mean, median, first and third quartiles of ESDs are reported. The results showed that in most cases, for each of the examinations, the individual ESD values are found to be comparable with, and higher than, those from Ghana and Tanzania, respectively. The mean ESD values are also found to be within the range of mean ESD values that have been previously been reported from countries outside Africa. When compared with the European Community (EC) reference values, the mean ESDs were found to be below the reference values in only two of the hospitals. The ranges found in this work are high and this indicates more attention needs to be given to X-ray facilities in the country. This also suggests that radiographic departments need to review their radiographic practices in order to bring their doses to optimum levels. Effective doses were also calculated from the ESD values. The mean effective doses were found to be generally low when compared with those found in the literature from other countries including two African countries. The radiographic parameters used for all the patients were also compared with the European criteria. It is recommended that the tube filtration at one hospital be increased. The importance of good regulatory activities and trained personnel is stressed in this work. Apart from the fact that the data provided in this work will be useful for the formulation of national guidance levels, they also provide patient dosimetry information on healthcare level IV countries.

Entities:  

Mesh:

Year:  2004        PMID: 15507418     DOI: 10.1259/bjr/55841517

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  4 in total

1.  Evaluation of dose-area product of common radiographic examinations towards establishing a preliminary diagnostic reference levels (PDRLs) in Southwestern Nigeria.

Authors:  Nnamdi N Jibiri; Christopher J Olowookere
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

2.  Survey of dose area product received by patients undergoing common radiological examinations in four centers in Nigeria.

Authors:  Bidemi I Akinlade; Idowu P Farai; Akintunde A Okunade
Journal:  J Appl Clin Med Phys       Date:  2012-07-05       Impact factor: 2.102

3.  Pediatric radiography entrance doses for some routine procedures in three hospitals within eastern Nigeria.

Authors:  N O Egbe; S O Inyang; O B Ibeagwa; N O Chiaghanam
Journal:  J Med Phys       Date:  2008-01

4.  Assessment of Pediatrics Radiation Dose from Routine X-Ray Examination at Jimma University Hospital, Southwest Ethiopia.

Authors:  Mesfin Zewdu; Elias Kadir; Melkamu Berhane
Journal:  Ethiop J Health Sci       Date:  2017-09
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.