The efficacy of the arrow staple device for securing central venous catheters to human skin.

Gregg K Motonaga1, Keith K Lee, Jeffrey R Kirsch.   

Abstract

We tested the hypothesis that the force necessary to dislodge a central venous catheter from human skin is related to its method of attachment. Specifically, we compared the peak axial force and torque required to remove a catheter hub attached to human skin with 2 3-0 silk sutures, 4 0.022-in. staples, or 4 0.025-in. staples. We used the tissue from the proximal end of discarded extremities after below-knee amputations to serve as a surrogate for the skin of living humans that would normally be the site for catheter stabilization. Central venous catheter hubs were secured to the area of grossly normal appearing tissue with 2 3-0 silk sutures or 4 staples. For each attachment method, 16 dislodgement trials were conducted. The mean peak force required for axial dislodgement with 3-0 silk sutures (40.9 +/- 10.7 N; n = 16) was more than that for 4 0.022-in. staples (34.0 +/- 7.2 N; n = 16;P = 0.04) but was not different from that for 4 0.025-in. staples (40.4 +/- 5.8 N; n = 16). The major finding of this study is that securing a central venous hub with 4 0.025-in. staples is as effective as 2 3-0 silk sutures in preventing axial dislodgement.

Entities:  

Mesh:

Year:  2004        PMID: 15502044     DOI: 10.1213/01.ANE.0000132544.15467.10

Source DB:  PubMed          Journal:  Anesth Analg        ISSN: 0003-2999            Impact factor:   5.108


  3 in total

1.  Blood and body fluid exposure related knowledge, attitude and practices of hospital based health care providers in United arab emirates.

Authors:  Moazzam Ali Zaidi; Robin Griffiths; Salem A Beshyah; Julie Myers; Mukarram A Zaidi
Journal:  Saf Health Work       Date:  2012-08-30

2.  Effectiveness of different central venous catheter fixation suture techniques: An in vitro crossover study.

Authors:  Manuel Florian Struck; Lars Friedrich; Stefan Schleifenbaum; Holger Kirsten; Wolfram Schummer; Bernd E Winkler
Journal:  PLoS One       Date:  2019-09-12       Impact factor: 3.240

3.  Catheter securement systems: comparison of two investigational devices to a sutureless securement device, a securement dressing, and sutures in a pig model.

Authors:  Laura F Rutledge; Daniel P DeCabooter; Shelley-Ann H Walters; Stéphanie F Bernatchez
Journal:  Intensive Care Med Exp       Date:  2015-08-27
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.