[The temporal behavior of the anchorage of a cement-free implanted acetabular cup in animal experiments].

H Bereiter1, M Bürgi, B A Rahn.   

Abstract

The mechanical qualities of the cementless press-fit cup of a hip prosthesis were examined in sheep. Primary anchorage of the cup is achieved by relative oversize of the cup in relation to the acetabulum, causing a press-fit effect. The surface of the cup consists of Sulmesh, a titanium net, which allows bony ingrowth. The quality of the mechanical anchorage in the acetabular bone was tested by measurements of the torque and cranio-caudal tilting moment needed to pull out the cup. It was shown that the secondary biological anchorage of the press-fit cup exceeds the primary mechanical anchorage (measured in previous in vitro experiments) by a factor of three. This improvement of the mechanical fixation of the press-fit cup in the acetabular bone can be explained by bony ingrowth into the Sulmesh and can be confirmed by histological findings.

Entities:  

Mesh:

Substances:

Year:  1992        PMID: 1549340

Source DB:  PubMed          Journal:  Orthopade        ISSN: 0085-4530            Impact factor:   1.087


  2 in total

Review 1.  [Uncemented arthroplasty of the hip].

Authors:  C von Schulze Pellengahr; A Fottner; S Utzschneider; M Schmitt-Sody; W Teske; T Lichtinger; S A Esenwein
Journal:  Orthopade       Date:  2009-05       Impact factor: 1.087

2.  Does osteoporosis reduce the primary tilting stability of cementless acetabular cups?

Authors:  Christoph von Schulze Pellengahr; Lars V von Engelhardt; Bernd Wegener; Peter E Müller; Andreas Fottner; Patrick Weber; Ole Ackermann; Matthias Lahner; Wolfram Teske
Journal:  BMC Musculoskelet Disord       Date:  2015-04-21       Impact factor: 2.362

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.