Literature DB >> 15473586

Pericarditis following permanent pacemaker insertion.

Yair Levy1, Ora Shovman, Chava Granit, David Luria, Osnat Gurevitz, David Bar-Lev, Michael Eldar, Yehuda Shoenfeld, Michael Glikson.   

Abstract

BACKGROUND: The appearance of pericarditis following insertion of a permanent pacemaker is not widely acknowledged in the literature.
OBJECTIVES: To describe our experience with pericarditis following 395 permanent pacemaker implantations over 2 years.
METHODS: We retrospectively reviewed the medical records of 395 consecutive patients in whom new pacing systems or pacemaker leads had been implanted over a 2 year period. We searched the records for pericarditis that developed within 1 month after pacemaker implantation according to the ICD-9 code. The incidence, clinical picture, response to treatment, and relationship to lead design and location were studied.
RESULTS: Eight cases (2%) of pericarditis following implantation were detected. Clinical manifestations in all patients were similar to those of post-pericardiotomy syndrome and included chest pain (n = 7), friction rub (n = 1), fever (n = 2), fatigue (n = 2), pleural effusion (n = 2), new atrial fibrillation (n = 2), elevated erythrocyte sedimentation rate (n = 4), and echcardiographic evidence of pericardial effusion (n = 8). All affected patients had undergone active fixation (screw-in) lead implantation in the atrial position. The incidence of pericarditis with screw-in atrial leads was 3% compared to 0% in other cases (P < 0.05).
CONCLUSIONS: Pericarditis is not uncommon following pacemaker implantation with active fixation atrial leads. Special attention should be paid to identifying pericardial complications following pacemaker implantation, especially when anticoagulant therapy is resumed or initiated. The use of passive fixation leads is likely to reduce the incidence of pericarditis but this issue should be further investigated.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15473586

Source DB:  PubMed          Journal:  Isr Med Assoc J            Impact factor:   0.892


  10 in total

Review 1.  Radiology of cardiac devices and their complications.

Authors:  J Dipoce; A Bernheim; H Spindola-Franco
Journal:  Br J Radiol       Date:  2014-11-20       Impact factor: 3.039

Review 2.  Post cardiac injury syndrome after initially uncomplicated CRT-D implantation: a case report and a systematic review.

Authors:  Farbod Sedaghat-Hamedani; Edgar Zitron; Elham Kayvanpour; Hanns-Martin Lorenz; Hugo A Katus; Benjamin Meder
Journal:  Clin Res Cardiol       Date:  2014-04-29       Impact factor: 5.460

3.  Post-cardiac injury syndrome following dual-chamber pacemaker placement.

Authors:  Joon Yong Moon; Mohammed Mogri
Journal:  Proc (Bayl Univ Med Cent)       Date:  2022-02-24

4.  Unusual presence of 'ghosts' following lead extraction for recurrent reactive pericarditis: a case report.

Authors:  Riccardo Bentivegna; Giuseppe Cattafi; Cristina Giannattasio; Antonella Moreo
Journal:  Eur Heart J Case Rep       Date:  2018-11-26

5.  A Rare Case of Acute Pleuropericarditis as a Complication of Permanent Pacemaker Insertion.

Authors:  Małgorzata Chlabicz; Piotr Jakim; Małgorzata Zalewska-Adamiec; Magdalena Róg-Makal; Sławomir Dobrzycki
Journal:  Am J Case Rep       Date:  2021-01-10

6.  Post cardiac injury syndrome successfully treated with medications: a report of two cases.

Authors:  Mu-Shiang Huang; Yan-Hua Su; Ju-Yi Chen
Journal:  BMC Cardiovasc Disord       Date:  2021-08-14       Impact factor: 2.298

7.  Pericarditis-induced hyponatremia after cardiac electronic implantable device (CEID) procedures.

Authors:  Elnaz Rakhshan; Seyed Abbas Mirabbasi; Bahar Khalighi; Koroush Khalighi
Journal:  Am J Case Rep       Date:  2015-04-26

8.  Unintended Harm and Benefit of the Implantable Defibrillator in an Unfortunate 19-Year-Old Male: Featuring a Sequence of Rare Life-threatening Complications of Cardiac Procedures.

Authors:  Amin Daoulah; Ali Ocheltree; Amir Lotfi; Sara Ocheltree; Alawi A Alsheikh-Ali; Abdul-Karim Al-Habib; Osama El-Sayed; Ali Haneef
Journal:  Indian Pacing Electrophysiol J       Date:  2013-08-01

9.  Alteplase Causing Cardiac Tamponade after Recent Cardiac Pacemaker Placement.

Authors:  Corey J Warf; Martin R Huecker; Daniel J O'Brien; Don A Bertolotti
Journal:  Clin Pract Cases Emerg Med       Date:  2018-05-18

10.  Pacemaker-Associated Post-cardiac Injury Syndrome Presenting with Tamponade and Recurrent Pleural Effusion.

Authors:  Young Ju Lee; Mahmood Mubasher; Abir Zainal; Tausif Syed; Mouhand F H Mohamed; Matthew Ferrantino; Ryan Hoefen
Journal:  Clin Med Insights Case Rep       Date:  2020-10-30
  10 in total