A method to remove artifacts in attenuation-corrected myocardial perfusion SPECT Introduced by misalignment between emission scan and CT-derived attenuation maps.

Harald Fricke1, Eva Fricke, Reiner Weise, Annett Kammeier, Oliver Lindner, Wolfgang Burchert.   

Abstract

UNLABELLED: Nonuniform soft-tissue attenuation affects the diagnostic accuracy of SPECT in myocardial perfusion imaging. The attenuation map required for attenuation correction can be acquired using x-ray tomography (CT). Frequent findings in attenuation-corrected images are defects in the apical and anterior myocardial wall. We assume that these are artifacts produced by misalignment of SPECT images and the attenuation map.
METHODS: One hundred forty patients underwent myocardial perfusion imaging with 99mTc-methoxyisobutylisonitrile. Twenty-seven of 140 showed pronounced defects in the apical or anterior wall only after CT-based attenuation correction. SPECT and corresponding CT slices were examined for misalignment in the ventrodorsal direction (y-direction) visually and by threshold-based delineation of the body surface. Mismatched studies were realigned and image reconstruction and analysis were redone. The effect of the correction was assessed visually and by semiquantitative analysis based on a 20-segment model using 4D-MSPECT.
RESULTS: In 15 of 27 patients, the improved coregistration led to smaller and less-pronounced defects in the regions mentioned. In 6 of 27 patients, former defects were judged as normal. No improvement was seen in only 4 patients. In these 4 subjects, the mismatch in the y-direction was <1 pixel (7 mm), and visual inspection suggested a coincident mismatch in the craniocaudal direction. In 2 cases, coregistration was not possible because the body outline extended beyond the CT field of view. Semiquantitative analysis revealed a significant increase of the relative uptake in the apex; in the apical segments of the anterior, septal, and inferior wall; and in the mid-anterior and mid-anteroseptal segment. Basal segments of the anterolateral, lateral, and inferolateral wall and the middle inferolateral segment showed a significant decrease of relative uptake.
CONCLUSION: Misalignment in the y-direction between SPECT and the attenuation map can lead to artifacts in the apical, septal, and anterior wall, which will appear as defects. It also can cause overcorrection in the basal inferior and lateral segments. There is evidence that mismatches along the other directions may have a similar effect. The coregistration of SPECT and the attenuation map needs to be verified for every patient, even when using integrated dual-modality imaging devices.

Entities:  

Mesh:

Year:  2004        PMID: 15471824

Source DB:  PubMed          Journal:  J Nucl Med        ISSN: 0161-5505            Impact factor:   10.057


  26 in total

1.  Impact of respiratory motion correction on SPECT myocardial perfusion imaging using a mechanically moving phantom assembly with variable cardiac defects.

Authors:  Irene Polycarpou; Isabelle Chrysanthou-Baustert; Ourania Demetriadou; Yiannis Parpottas; Christoforos Panagidis; Paul K Marsden; Lefteris Livieratos
Journal:  J Nucl Cardiol       Date:  2015-12-16       Impact factor: 5.952

2.  Use of coronary calcium score scans from stand-alone multislice computed tomography for attenuation correction of myocardial perfusion SPECT.

Authors:  Tiziano Schepis; Oliver Gaemperli; Pascal Koepfli; Christine Rüegg; Cyrill Burger; Sebastian Leschka; Lotus Desbiolles; Lars Husmann; Hatem Alkadhi; Philipp A Kaufmann
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-08-04       Impact factor: 9.236

3.  Myocardial perfusion and function single photon emission computed tomography.

Authors:  Christopher L Hansen; Richard A Goldstein; Daniel S Berman; Keith B Churchwell; C David Cooke; James R Corbett; S James Cullom; Seth T Dahlberg; James R Galt; Ravi K Garg; Gary V Heller; Mark C Hyun; Lynne L Johnson; April Mann; Benjamin D McCallister; Raymond Taillefer; R Parker Ward; John J Mahmarian
Journal:  J Nucl Cardiol       Date:  2006-11       Impact factor: 5.952

4.  Attenuation correction in cardiac SPECT: the boy who cried wolf?

Authors:  Guido Germano; Piotr J Slomka; Daniel S Berman
Journal:  J Nucl Cardiol       Date:  2007-01       Impact factor: 5.952

5.  How serious a problem for myocardial perfusion assessment is moderate misregistration between SPECT and CT?

Authors:  Kenneth J Nichols
Journal:  J Nucl Cardiol       Date:  2007-04       Impact factor: 5.952

6.  Prevalence of misregistration between SPECT and CT for attenuation-corrected myocardial perfusion SPECT.

Authors:  Sibyll Goetze; Richard L Wahl
Journal:  J Nucl Cardiol       Date:  2007-04       Impact factor: 5.952

Review 7.  Myocardial perfusion and function: single photon emission computed tomography.

Authors:  Christopher L Hansen; Richard A Goldstein; Olakunle O Akinboboye; Daniel S Berman; Elias H Botvinick; Keith B Churchwell; C David Cooke; James R Corbett; S James Cullom; Seth T Dahlberg; Regina S Druz; Edward P Ficaro; James R Galt; Ravi K Garg; Guido Germano; Gary V Heller; Milena J Henzlova; Mark C Hyun; Lynne L Johnson; April Mann; Benjamin D McCallister; Robert A Quaife; Terrence D Ruddy; Senthil N Sundaram; Raymond Taillefer; R Parker Ward; John J Mahmarian
Journal:  J Nucl Cardiol       Date:  2007 Nov-Dec       Impact factor: 5.952

8.  Automated quality control of emission-transmission misalignment for attenuation correction in myocardial perfusion imaging with SPECT-CT systems.

Authors:  Ji Chen; Serpil F Caputlu-Wilson; Hongcheng Shi; James R Galt; Tracy L Faber; Ernest V Garcia
Journal:  J Nucl Cardiol       Date:  2006 Jan-Feb       Impact factor: 5.952

9.  Use of MRI to assess the prediction of heart motion with gross body motion in myocardial perfusion imaging by stereotracking of markers on the body surface.

Authors:  Michael A King; Joyoni Dey; Karen Johnson; Paul Dasari; Joyeeta M Mukherjee; Joseph E McNamara; Arda Konik; Cliff Lindsay; Shaokuan Zheng; Dennis Coughlin
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

10.  Myocardial perfusion imaging and coronary calcium scoring with a two-slice SPECT/CT system: can the attenuation map be calculated from the calcium scoring CT scan?

Authors:  Christian Wenning; Kambiz Rahbar; Alexis Vrachimis; Otmar Schober; Michael Schäfers; Lars Stegger
Journal:  Eur J Nucl Med Mol Imaging       Date:  2013-02-22       Impact factor: 9.236

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.