Literature DB >> 15469863

Evaluation of Advia Centaur automated chemiluminescence immunoassay for determining total homocysteine in plasma.

Karine Demuth1, Véronique Ducros, Samantha Michelsohn, Jean-Louis Paul.   

Abstract

BACKGROUND: A lot of methods are now available for total plasma homocysteine (tHcy) determination. HPLC with fluorescence detection were the most widely used methods until recently but immunoassays, easier to use, currently supplant in-house laboratory methods.
METHODS: The latest commercial automated chemiluminescence immunoassay (ICL) for tHcy, developed on the ADVIA Centaur analyzer (Bayer Diagnostics), was fully evaluated and compared with a current HPLC method.
RESULTS: The ICL-ADVIA Centaur method had a detection limit <0.9 micromol/l and was linear for tHcy between 2.4 and 58.8 micromol/l. The within- and between assay imprecision was <6% and <7%, respectively. The analytical recovery ranged from 93.5% to 109.7%. The comparison of 168 clinical plasma specimens indicated a good correlation between the two methods (r=0.96, p<0.0001) but a systematic positive bias for ICL-ADVIA Centaur method (mean difference=3.0 micromol/l).
CONCLUSIONS: The ADVIA Centaur method is a useful and practicable alternative to HPLC methods for tHcy determination. The ICL method is even more convenient than chromatographic methods for routine use because of its significantly higher throughput (200 samples/h vs. 40 per 24 h). However, the agreement among methods is insufficient to allow them to be used interchangeably.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15469863     DOI: 10.1016/j.cccn.2004.06.012

Source DB:  PubMed          Journal:  Clin Chim Acta        ISSN: 0009-8981            Impact factor:   3.786


  4 in total

1.  Connexin30 deficiency causes instrastrial fluid-blood barrier disruption within the cochlear stria vascularis.

Authors:  Martine Cohen-Salmon; Béatrice Regnault; Nadège Cayet; Dorothée Caille; Karine Demuth; Jean-Pierre Hardelin; Nathalie Janel; Paolo Meda; Christine Petit
Journal:  Proc Natl Acad Sci U S A       Date:  2007-03-30       Impact factor: 11.205

2.  Levels of plasma homocysteine in pseudoexfoliation glaucoma.

Authors:  Laura Tranchina; Marco Centofanti; Francesco Oddone; Lucia Tanga; Gloria Roberti; Laura Liberatoscioli; Claudio Cortese; Gianluca Manni
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2010-08-26       Impact factor: 3.117

3.  Determination of Autosomal Dominant or Recessive Methionine Adenosyltransferase I/III Deficiencies Based on Clinical and Molecular Studies.

Authors:  Yoo-Mi Kim; Ja Hye Kim; Jin Choi; Kim Gu-Hwan; Jae-Min Kim; Minji Kang; In-Hee Choi; Chong Kun Cheon; Young Bae Sohn; Marco Maccarana; Han-Wook Yoo; Beom Hee Lee
Journal:  Mol Med       Date:  2016-02-18       Impact factor: 6.354

4.  Analysis of serum homocysteine in the laboratory practice - comparison of the direct chemiluminescence immunoassay and high performance liquid chromatography coupled with fluorescent detection.

Authors:  Łukasz Paprotny; Dorota Wianowska; Magdalena Izdebska; Agnieszka Celejewska; Dorota Szewczak; Janusz Solski
Journal:  Biochem Med (Zagreb)       Date:  2020-08-05       Impact factor: 2.313