Literature DB >> 15449120

Clinical accuracy evaluation of femoral canal preparation using the ROBODOC system.

Shunsaku Nishihara1, Nobuhiko Sugano, Takashi Nishii, Hisashi Tanaka, Nobuo Nakamura, Hideki Yoshikawa, Takahiro Ochi.   

Abstract

The purpose of this study was to evaluate the clinical accuracy of femoral canal preparation using postoperative reconstructed computed tomography (CT) images of 75 consecutive total hip arthroplasties performed with a two-pin-based ROBODOC system. Intraoperatively, the robot milled the femoral canal according to the preoperative planning performed with preoperative CT data and the ROBODOC workstation. Postoperative CT data was obtained 1 month postoperatively. Anteroposterior and lateral synthetic radiographs and axial images were reconstructed from CT data on the workstation. The mean difference between the preoperative planning and the postoperative CT images was less than 5% in canal fill, less than 1 mm in gap, and less than 1 degrees in the mediolateral and anteroposterior alignment. Clinical femoral canal preparation using the ROBODOC system results in a high degree of accuracy.

Entities:  

Mesh:

Year:  2004        PMID: 15449120     DOI: 10.1007/s00776-004-0804-5

Source DB:  PubMed          Journal:  J Orthop Sci        ISSN: 0949-2658            Impact factor:   1.601


  6 in total

1.  Level of osteotomy is relevant to obtain better union and clinical results in patients with severe hip dysplasia operated on with total hip arthroplasty and shortening osteotomy using a cementless, rectangular femoral component.

Authors:  Mahmut Enes Kayaalp; Ata Can; Fahri Erdogan; Mahmut Kursat Ozsahin; Onder Aydingoz; Gokhan Kaynak
Journal:  Arch Orthop Trauma Surg       Date:  2020-11-08       Impact factor: 3.067

Review 2.  Robotics in trauma and orthopaedics.

Authors:  Karthik Karuppiah; Joydeep Sinha
Journal:  Ann R Coll Surg Engl       Date:  2018-05       Impact factor: 1.891

3.  An experimental evaluation of the force requirements for robotic mastoidectomy.

Authors:  Neal P Dillon; Louis B Kratchman; Mary S Dietrich; Robert F Labadie; Robert J Webster; Thomas J Withrow
Journal:  Otol Neurotol       Date:  2013-09       Impact factor: 2.311

4.  Sensing small interaction forces through proprioception.

Authors:  Fazlur Rashid; Devin Burns; Yun Seong Song
Journal:  Sci Rep       Date:  2021-11-08       Impact factor: 4.379

Review 5.  Computer-assisted orthopaedic surgery and robotic surgery in total hip arthroplasty.

Authors:  Nobuhiko Sugano
Journal:  Clin Orthop Surg       Date:  2013-02-20

Review 6.  A history of robots: from science fiction to surgical robotics.

Authors:  N G Hockstein; C G Gourin; R A Faust; D J Terris
Journal:  J Robot Surg       Date:  2007-03-17
  6 in total