Literature DB >> 1543157

Evaluation of four automated hematology analyzers. A comparative study of differential counts (imprecision and inaccuracy).

M Buttarello1, M Gadotti, C Lorenz, E Toffalori, N Ceschini, A Valentini, P Rizzotti.   

Abstract

The authors evaluated the performance of four modern, commercially available hematology analyzers for imprecision and inaccuracy in determining the leukocyte differential count. The evaluation was performed according to International Committee for Standardization in Haematology protocols and the National Committee for Clinical Laboratory Standards H20-T standard, using the same group of patients simultaneously. Imprecision was very low among all the analyzers for neutrophils and lymphocytes (coefficient of variation maximum = 4.12%), whereas for the other leukocyte populations it tended to increase as their presence percentage decreased. The imprecision of the analyzers was still lower than that of the microscopic method. The correlation with the manual 800 cell count (inaccuracy) was good for neutrophils, lymphocytes, and eosinophils (r = 0.974 to 0.888), less so for monocytes (r = 0.757 to 0.490), whereas it was poor for basophils (r = 0.532 to 0.078).

Entities:  

Mesh:

Year:  1992        PMID: 1543157     DOI: 10.1093/ajcp/97.3.345

Source DB:  PubMed          Journal:  Am J Clin Pathol        ISSN: 0002-9173            Impact factor:   2.493


  4 in total

1.  Validity of Selected WBC Differentiation Flags in Sysmex XT-1800i.

Authors:  Parya Bameni Moghaddam; Fatemeh Mahjoub; Amirhossein Emami; Alireza Abdollahi
Journal:  Iran J Pathol       Date:  2016

2.  The white cell differential: personal observations.

Authors:  D M Reardon; B Warner
Journal:  J Clin Pathol       Date:  1993-04       Impact factor: 3.411

3.  Blood film in the era of streaming cells.

Authors:  Diego Villa Clé
Journal:  Rev Bras Hematol Hemoter       Date:  2017-08-23

4.  Mediation Effect of Neutrophil Lymphocyte Ratio on Cardiometabolic Risk Factors and Cardiovascular Events.

Authors:  Teeranan Angkananard; Thunyarat Anothaisintawee; Atiporn Ingsathit; Mark McEvoy; Kongpop Silapat; John Attia; Piyamitr Sritara; Ammarin Thakkinstian
Journal:  Sci Rep       Date:  2019-02-22       Impact factor: 4.379